跳到正文
原文
Nature 健康护理主题·· 2026-10-13AI 评分35

AnchorDoc低剖面缝线锚定装置用于非血管导管机械稳定性的研究

Mechanical stabilization of non-vascular catheters using a low-profile suture-anchored device

AI 导读

AnchorDoc在干燥(10.2Fr:31.1±1.4N)和湿润条件下峰值抗滑脱力均显著高于StatLock(20.5±2.4N)和Roman Sandal(18.8±2.3N,p=0.05),动态步态模拟显示其失效时间延长3.2倍,且静态流量测试未发现导管腔道梗阻。

正文

Scientific Reports (2026) Cite this article

We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.

Abstract

To evaluate the mechanical performance of AnchorDoc, a novel, low-profile, two-piece, suture-anchored device, against standard securement methods. We hypothesized that AnchorDoc would provide superior mechanical stability without compromising catheter patency. Using a custom anatomically-informed gelatin phantom of the human flank with porcine-skin overlay, we compared the performance of AnchorDoc against a standard adhesive device (StatLock) and a suture-tape technique (Roman sandal) for 8.5, 10.2, and 12 French (Fr) catheters. We conducted: (1) static tensile testing under dry and wet conditions to measure peak force to catheter slip (≥ 5 mm); (2) dynamic gait simulation to assess time-to-failure; and (3) static flow testing to evaluate for lumen compromise. In tensile testing, AnchorDoc required a significantly higher dislodgement force than both StatLock and Roman Sandal under dry (e.g., 10.2 Fr: 31.1 ± 1.4 N vs. 20.5 ± 2.4 N and 18.8 ± 2.3 N, respectively; p < 0.0001) and wet conditions (25.5 ± 1.0 N vs. 10.9 ± 1.3 N and 18.7 ± 1.0 N; p < 0.0001). The adhesive device’s retention strength degraded by nearly 50% when wet, while AnchorDoc maintained its superiority. In dynamic gait simulation, Kaplan-Meier analysis showed that AnchorDoc had a significantly higher survival probability (fewer failures) while tested over 12,000 cycles compared to both comparators (p < 0.05). Flow testing confirmed that AnchorDoc did not cause a statistically significant reduction in flow rate or lumen diameter across all catheter sizes (p > 0.05). The AnchorDoc device provides significantly greater resistance to both static and dynamic dislodgement forces than common adhesive and suture-tape techniques, particularly in a laboratory-simulated wet environment, without impairing catheter function. This suture-anchored, two-part platform represents a promising solution to mitigate the risk of non-vascular catheter dislodgement.

Abbreviations

ANOVA:

Analysis of variance

CAD:

Computer-aided design

CI:

Confidence interval

CIRSE:

Cardiovascular and interventional radiological society of europe

Fr:

French (catheter gauge)

IR:

Interventional radiology

IQR:

Interquartile range

kPa:

Kilopascal

lbf:

Pound-force

N:

Newton

NS:

Normal saline (0.9% NaCl)

PLA:

Polylactic acid

SIR:

Society of interventional radiology

SOP:

Standard operating procedure

US:

Ultrasound

UTM:

Universal testing machine

Funding

Supported by the National Center for Advancing Translational Sciences of the National Institutes of Health under Award Number UL1TR002378.

Author information

Author notes

  1. These authors contributed equally to this work: Shreyas Naidu and Chang Zhou.

Authors and Affiliations

  1. Division of Interventional Radiology and Image-Guided Medicine, Department of Radiology and Imaging Science, Emory University School of Medicine, Atlanta, Georgia

    Shreyas Naidu, Zachary Bercu, Hanzhou Li & John Moon

  2. Division of Vascular and Interventional Radiology, University of Florida College of Medicine, Gainesville, FL, USA

    Shreyas Naidu

  3. George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, 801 Ferst Dr, Atlanta, GA, USA

    Chang Zhou & Qiyue Ma

  4. The Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology, 313 Ferst Dr, Atlanta, GA, USA

    Yue Chen

Authors

  1. Shreyas Naidu
  2. Chang Zhou
  3. Qiyue Ma
  4. Zachary Bercu
  5. Hanzhou Li
  6. John Moon
  7. Yue Chen

Corresponding author

Correspondence to Yue Chen.

Ethics declarations

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Naidu, S., Zhou, C., Ma, Q. et al. Mechanical stabilization of non-vascular catheters using a low-profile suture-anchored device. Sci Rep (2026). https://doi.org/10.1038/s41598-026-68117-1

Download citation

  • Received: 02 November 2025

  • Accepted: 19 August 2026

  • Published: 13 October 2026

  • DOI: https://doi.org/10.1038/s41598-026-68117-1

Keywords

来源:Nature 健康护理主题 · nature.com