通过心血管硬件模拟器的振荡传递函数分析非侵入性估计肱动脉压力波形
Non-invasive estimation of brachial pressure waveforms via oscillometric transfer function analysis using cardiovascular hardware simulator
研究使用心血管硬件模拟器和六种上臂仿真配置,重建标准化肱动脉压力波形。结果显示,个体传递函数(ITFs)重建的皮尔逊相关系数为0.9865,归一化均方根误差为0.0528,而留一仿真配置外推的通用传递函数(GTFs)分别为0.9725和0.0713。这些结果支持在测试模拟器中重建标准化波形,但未证实系统线性或绝对血压准确性。
- Article
- Open access
- Published: 11 November 2026
Scientific Reports (2026) Cite this article
We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.
Abstract
Cuff oscillations are dynamically distorted relative to arterial pressure. We investigated reconstruction of normalized brachial arterial pressure waveforms using a cardiovascular hardware simulator with six upper-arm phantom configurations. Transfer functions were identified from random excitation and evaluated on separate pulsatile recordings at 60–96 beats/min and nominal cuff pressures of 40, 100 and 160 mmHg. Eighteen individual transfer functions (ITFs) were estimated. Generalized transfer functions (GTFs) were constructed by equally weighted complex averaging; leave-one-phantom-configuration-out evaluation excluded the test configuration. On the same 72 conditions, ITF reconstruction yielded Pearson correlation r = 0.9865 ± 0.0102 and normalized root-mean-square error (nRMSE) = 0.0528 ± 0.0213, whereas held-out GTF yielded r = 0.9725 ± 0.0192 and nRMSE = 0.0713 ± 0.0235 (mean ± standard deviation). A phenomenological standard-linear-solid model fitted the complex responses with R2 = 0.9628 ± 0.0161. Exploratory analysis of peak-height ratios identified residual reconstruction errors and sensitivity to peak-detection settings. These results support normalized waveform reconstruction in the tested simulator; they do not establish system linearity, absolute blood-pressure accuracy or clinical generalizability.
Materials availability
Not applicable.
Funding
This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korea government (MSIT) (No. RS-2024-00350917).
Author information
Authors and Affiliations
Department of Mechanical Engineering, Korea Advanced Institute of Science and Technology (KAIST), Daejeon, 34141, Republic of Korea
Junki Hong, Bomi Lee & Yong-Hwa Park
Department of Electronic Engineering, Korea National University of Transportation, Chungju-si, Chungcheongbuk-do, Republic of Korea
Jae-Hak Jeong
Authors
- Junki Hong
- Bomi Lee
- Jae-Hak Jeong
- Yong-Hwa Park
Corresponding author
Correspondence to Yong-Hwa Park.
Ethics declarations
Competing interests
The authors declare no competing interests.
Ethics approval and consent to participate
This study used a cardiovascular hardware simulator and did not involve human participants or animal subjects. Ethics approval and informed consent were not applicable.
Consent for publication
Not applicable.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Hong, J., Lee, B., Jeong, JH. et al. Non-invasive estimation of brachial pressure waveforms via oscillometric transfer function analysis using cardiovascular hardware simulator. Sci Rep (2026). https://doi.org/10.1038/s41598-026-79703-8
Received: 25 February 2026
Accepted: 03 November 2026
Published: 11 November 2026
DOI: https://doi.org/10.1038/s41598-026-79703-8
Keywords
来源:Nature 健康护理主题 · nature.com