跳到正文
原文
Nature 健康护理主题·· 6 天前AI 评分35

辨别转移但校准不转移:糖尿病视网膜病变筛查相机研究

Discrimination transfers but calibration does not across fundus cameras in diabetic retinopathy screening

AI 导读

研究发现,用于糖尿病视网膜病变(DR)筛查的深度学习模型在不同相机之间保持了辨别能力,但校准效果未保持一致。在手持相机上,未经校准的模型校准误差显著增加;通过局部标签校准可恢复校准效果。新相机需重新校准以确保准确性。

正文

Scientific Reports (2026) Cite this article

We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.

Abstract

Deep-learning screening for referable diabetic retinopathy (DR) increasingly runs on cameras it was not developed on. We asked whether a referable-DR classifier trained on tabletop fundus photographs (BRSET) keeps its calibration — the agreement between predicted probability and observed frequency — as well as its discrimination when applied without adaptation to handheld images (mBRSET) and an independent external dataset (APTOS 2019). Across two networks and five training runs, zero-shot discrimination stayed within 0.03–0.04 AUROC of networks trained on each target (AUROC 0.89–0.96). Calibration did not: on handheld images, expected calibration error (ECE) was 0.06–0.07 without recalibration and 0.10–0.11 with the source-fitted calibrator (0.009 in-domain), and a threshold giving 90% sensitivity on the source detected only 42–47% of referable handheld images. On the handheld target, prevalence matching attributed about four-fifths (\(79\pm 10\%\)) of the calibration loss to its higher disease prevalence; on the external target, prevalence and other dataset differences partly offset each other. Prevalence correction without labels did not restore calibration, whereas Platt or isotonic recalibration with 100–200 labelled target images did (ECE 0.03–0.04). Half-precision models preserved discrimination and calibration; 8-bit integer quantization reduced zero-shot AUROC by 0.07–0.14. New cameras therefore require recalibration on local labels.

Funding

The author(s) received no specific funding for this work.

Author information

Authors and Affiliations

  1. Independent Researcher, San Jose, California, 95126, United States of America

    Sairam Tabibu & Mugdha Abhyankar

Authors

  1. Sairam Tabibu
  2. Mugdha Abhyankar

Corresponding author

Correspondence to Sairam Tabibu.

Ethics declarations

Competing interests

The authors declare no competing interests.

Additional information

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Tabibu, S., Abhyankar, M. Discrimination transfers but calibration does not across fundus cameras in diabetic retinopathy screening. Sci Rep (2026). https://doi.org/10.1038/s41598-026-78618-8

Download citation

  • Received: 11 July 2026

  • Accepted: 27 October 2026

  • Published: 14 November 2026

  • DOI: https://doi.org/10.1038/s41598-026-78618-8

Keywords

来源:Nature 健康护理主题 · nature.com