三维向量数字化工作流程评估根管治疗精度的体外研究:巴西某大学团队验证

Accuracy of guided endodontic access evaluated through a fully digital workflow with three-dimensional vector-based measurement: an in vitro study

AI 导读

巴西某大学团队验证三维向量数字化工作流程在体外模型中评估根管治疗精度的可行性,使用18个标准化树脂模型,平均偏差0.260±0.221 mm(95%CI 0.150-0.369 mm),88.9%样本偏差≤0.5 mm(符合临床阈值),ICC 3,1=0.998(可靠性极好),p<0.001(统计学显著)。

正文

Scientific Reports (2026) Cite this article

We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.

Abstract

To characterize a fully digital, vector-based workflow for assessing the accuracy of static guided endodontic access and to determine the precision of access execution under controlled laboratory conditions, minimizing methodological bias associated with tomographic-based measurements. This in vitro experimental study used 18 three-dimensionally printed resin replicas of a standardized maxillary central incisor (tooth #11) model with a flattened apical surface allowing direct visual verification of the access point. Static guided endodontic access was performed by a single calibrated operator using a static guide planned in Implant Studio (3Shape, Copenhagen, Denmark) with a metal sleeve (1.5-mm internal diameter; 7.5-mm length). The dependent variable was the linear deviation (mm) between the planned and executed access centers, obtained by superimposing pre- and post-operative STL meshes in Medit Design software. Intra-operator reliability was assessed according to COSMIN standards (ICC3,1, standard error of measurement [SEM], minimal detectable change at 95% confidence [MDC95], and Bland–Altman analysis). The mean linear deviation was 0.260 ± 0.221 mm (95% CI 0.150–0.369 mm; median 0.249 mm). Bland–Altman analysis revealed a bias of 0.004 mm and narrow limits of agreement (− 0.024 to 0.032 mm). Intra-operator reliability was excellent (ICC3,1= 0.998; SEM = 0.0101 mm; MDC95 = 0.028 mm). Among the included samples, 88.9% showed a deviation ≤ 0.5 mm, all remained below the 0.7 mm inclusion threshold by definition. The deviation was statistically greater than zero (Wilcoxon signed-rank, p < 0.001), but a sensitivity analysis excluding the two largest observations confirmed the robustness of the findings. Under controlled laboratory conditions, static guided endodontic access demonstrated high accuracy, with deviations predominantly within clinically accepted limits. The proposed fully digital, vector-based workflow proved to be a precise and highly reproducible alternative to tomographic-based measurement methods for assessing procedural accuracy in experimental endodontic research.

Funding

No funding was received for this manuscript.

Author information

Authors and Affiliations

  1. Western Paraná State University (UNIOESTE), Cascavel, PR, 85819-110, Brazil

    André Luis Dias Carvalho

  2. Graduate Program in Integrated Dentistry, State University of Maringá (UEM), Maringá, PR, Brazil

    Elen Mariane Dal Bosco Klein, Diogo Henrique Nakaie, Isabela Inoue Kussaba & Eduardo Kurihara

Authors

  1. Elen Mariane Dal Bosco Klein
  2. André Luis Dias Carvalho
  3. Diogo Henrique Nakaie
  4. Isabela Inoue Kussaba
  5. Eduardo Kurihara

Corresponding author

Correspondence to André Luis Dias Carvalho.

Ethics declarations

Competing interests

The authors declare no competing interests.

Ethical approval

This in vitro study used standardized synthetic models without human participants, records, or biological materials; thus, institutional review board approval was not required.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Klein, E.M.D.B., Carvalho, A.L.D., Nakaie, D.H. et al. Accuracy of guided endodontic access evaluated through a fully digital workflow with three-dimensional vector-based measurement: an in vitro study. Sci Rep (2026). https://doi.org/10.1038/s41598-026-69828-1

Download citation

  • Received: 13 July 2026

  • Accepted: 31 August 2026

  • Published: 26 October 2026

  • DOI: https://doi.org/10.1038/s41598-026-69828-1