Feasibility assessment of novel binocular eye alignment monitor device for intermittent exotropia
新型双眼眼位对齐监测仪 BEAM 在 22 名双眼正常视力者和 15 名间歇性外斜视患者中完成可行性评估,可在 90 分钟观影期间客观记录眼位,时间分辨率 8 毫秒。两组偏离时间百分比及最大、平均、中位偏斜量差异显著(p<0.01),临床医生测得的最大偏斜量与 BEAM 显著相关(p<0.03)。研究团队认为仍需随机临床试验进一步验证。
- Article
- Open access
- Published: 27 October 2026
Scientific Reports (2026) Cite this article
We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.
Abstract
Intermittent exotropia exhibits inherent variability in the percentage of time the eye deviates and the amount of exodeviation; therefore, assessing the magnitude and duration of the eye’s deviation is challenging. A portable monitoring device for longer assessments can enhance evaluation and management, providing an objective tool for assessing therapeutic rehabilitative interventions. The Binocular Eye Alignment Monitor (BEAM) was developed to objectively evaluate eye alignment over an extended period. Eye position was objectively recorded from 22 participants with binocularly normal vision and 15 with intermittent exotropia. Participants watched an age-appropriate movie of their choice on a television located 6 m away for 90 min after undergoing a strabismus evaluation in two sessions, each occurring on a different day. Test and retest sessions were analyzed for the percentage of test time deviated, as well as the maximum, mean, and median deviation magnitudes. Repeatability was assessed using the intraclass correlation coefficient and Bland-Altman plots. The BEAM has a temporal resolution of 8 milliseconds and a conservative spatial resolution of 2Δ. A Mann-Whitney U-test revealed significant differences between groups for the percentage of time deviated and the maximum, mean, and median deviation (p < 0.01). Bland-Altman plots indicated good repeatability and narrow limits of agreement in the binocular normal vision group, whereas the intermittent exotropia group showed greater variability. Linear regression analysis showed a significant correlation between the maximum deviation measured by a clinician and the BEAM (p < 0.03). These feasibility results warrant further evaluation of the BEAM for future randomized clinical trials.
Acknowledgements
The co-authors appreciate the participants’ time in participating in our study.
Funding
National Eye Institute of the National Institutes of Health (NEI R01EY023261) to TLA.
Author information
Authors and Affiliations
Department of Biomedical Engineering, New Jersey Institute of Technology, Newark, NJ, USA
Christopher J. Morris & Tara L. Alvarez
Pennsylvania College of Optometry, Salus at Drexel University, Elkins Park, PA, USA
Erin Jenewein & Mitchell Scheiman
Authors
- Christopher J. Morris
- Erin Jenewein
- Mitchell Scheiman
- Tara L. Alvarez
Corresponding author
Correspondence to Tara L. Alvarez.
Ethics declarations
Consent for publication
All co-authors and participants gave informed consent to the publication of the enclosed research.
Competing interests
The authors declare no competing interests.
Disclosures
CJ Morris, TL Alvarez, and M Scheiman are co-inventors on patent application US20240074699A1, System and Method for Quantification and Feedback of Eye Deviations. The co-authors, given their research responsibilities, do not intend to pursue commercialization of this device.
Ethics approval and informed consent to participate
The experimental protocol was approved by the Institutional Review Boards of New Jersey Institute of Technology and Drexel University, in accordance with the Declaration of Helsinki. Adolescents/children provided written assent with the written informed consent of a parent or legal guardian. Adults gave written informed consent.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Morris, C.J., Jenewein, E., Scheiman, M. et al. Feasibility assessment of novel binocular eye alignment monitor device for intermittent exotropia. Sci Rep (2026). https://doi.org/10.1038/s41598-026-76743-y
Received: 28 January 2026
Accepted: 14 October 2026
Published: 27 October 2026
DOI: https://doi.org/10.1038/s41598-026-76743-y
来源:Nature 健康护理主题 · nature.com