传感器皮肤接触压力作为可穿戴光电容积脉搏波传感的隐藏混杂因素

Sensor-skin contact pressure as a hidden confounder in wearable photoplethysmography biosensing

AI 导读

该研究系统性综述 29 项涉及光电容积脉搏波(PPG)采集与袖带无创血压估算的实验证据,指出接触压力不足会降低光学稳定性并增加运动伪影,压力过大会压缩浅表血管、衰减收缩峰并抑制重搏切迹;由于各研究采用的统计指标不一致,无法对不同信号处理方法进行量化排序。研究还发现目前接触压力的测量、控制与报告存在显著的方法学不一致,并提出标准化压力感知实践以提升可重复性与临床转化。

正文

npj Biosensing (2026) Cite this article

We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.

Abstract

Photoplethysmography (PPG) is increasingly used for cuffless blood pressure (BP) monitoring and wearable cardiovascular sensing. However, the reliability of PPG-derived digital biomarkers is highly sensitive to sensor–skin contact pressure, which influences optical coupling, vascular compression, and motion susceptibility. Despite its importance, contact pressure is often poorly controlled or insufficiently reported, introducing a hidden source of variability that can compromise signal integrity, downstream model performance, and the reproducibility of wearable BP systems. Here, we systematically review experimental evidence on the influence of contact pressure in PPG-based cuffless BP estimation. Across 29 studies of PPG acquisition, BP estimation, and related pulse sensing, those that examined contact loading reported the following qualitative patterns: insufficient pressure reduced optical stability and increased motion artefacts, whereas excessive pressure compressed superficial vasculature, attenuated systolic peaks, and suppressed the dicrotic notch. Within studies that varied contact loading, favourable conditions depended on the anatomical site, sensor, and outcome; better waveform quality did not by itself establish lower BP-estimation error or a universal pressure target. Low, moderate, and high loading conditions are therefore used only as study-defined descriptions of signal-level effects. Because reported performance metrics represented different statistical constructs, outcomes were summarized separately for signal quality, waveform morphology, timing features, and BP-estimation error; the evidence does not support quantitative ranking of classical signal-processing, statistical, and machine-learning approaches. We further identify substantial methodological inconsistency in pressure measurement, control, and reporting, and propose standardized pressure-aware practices to improve reproducibility, benchmarking, device design, and the clinical translation of cuffless BP monitoring.

Funding

This work was supported by Khalifa University through the Faculty Start-Up Grant (FSU-2025-001). The funder had no role in the study design, data collection, analysis, interpretation, or preparation of the manuscript.

Author information

Author notes

  1. These authors contributed equally: Mohamed Elgendi, Azza Elkhalifa, Anood Alali.

Authors and Affiliations

  1. Department of Biomedical Engineering and Biotechnology, Khalifa University, Abu Dhabi, United Arab Emirates

    Mohamed Elgendi, Azza Elkhalifa, Anood Alali, Sara Alameri, Shahad Alzaabi & Najoud Alhammadi

  2. Healthcare Engineering Innovation Group, Khalifa University, Abu Dhabi, United Arab Emirates

    Mohamed Elgendi

  3. Translational and Clinical Research Institute, Faculty of Medical Sciences, Newcastle University, Newcastle upon Tyne, UK

    John Allen

  4. Department of Mechanical Engineering, Massachusetts Institute of Technology (MIT), Cambridge, MA, USA

    Richard Ribon Fletcher

Authors

  1. Mohamed Elgendi
  2. Azza Elkhalifa
  3. Anood Alali
  4. Sara Alameri
  5. Shahad Alzaabi
  6. Najoud Alhammadi
  7. John Allen
  8. Richard Ribon Fletcher

Corresponding author

Correspondence to Mohamed Elgendi.

Ethics declarations

Competing interests

M.E. is an Editorial Board Member of npj Biosensing and Communications Medicine. M.E. was not involved in the journal’s review of, or decisions related to, this manuscript. The other authors declare no competing financial or non-financial interests.

Additional information

Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Elgendi, M., Elkhalifa, A., Alali, A. et al. Sensor-skin contact pressure as a hidden confounder in wearable photoplethysmography biosensing. npj Biosensing (2026). https://doi.org/10.1038/s44328-026-00120-y

Download citation

  • Received: 27 May 2026

  • Accepted: 14 October 2026

  • Published: 15 November 2026

  • DOI: https://doi.org/10.1038/s44328-026-00120-y

来源:Nature 健康护理主题 · nature.com