A key player in the cuffless blood pressure devices market has exited – a cautionary tale for cuffless blood pressure monitoring?

AI 导读

新加坡HealthSTATS International公司宣布终止运营,其早期商用无袖带血压监测设备BPro停止支持,相关监管批准已撤回。BPro曾通过ESH协议及ANSI/AAMI SP10:2002验证,并在新加坡、美国、欧洲和澳大利亚获得医疗器械监管批准,用于动态血压监测。

正文
  • Correspondence
  • Published: 06 October 2026

Journal of Human Hypertension (2026) Cite this article

The pursuit of reliable and convenient cuffless blood pressure (BP) monitoring has gained momentum both in the scientific community and commercial devices markets. Yet, amongst the flurry of activity, a once key player in the cuffless BP space quietly withdrew after almost two decades without causing so much as a ripple.

BPro (HealthSTATS International, Singapore), one of the earliest commercial cuffless BP devices, is a watch-like, wrist wearable, cuff-calibrated device that obtained pulse waveform signals using tonometry to estimate BP. Having passed the European Society of Hypertension (ESH) protocol and American National Standard ANSI/AAMI SP10:2002 validation criteria for cuff-based sphygmomanometers [1]— before any validation standards specific for cuffless BP devices were established—BPro subsequently gained medical device regulatory approvals (including in Singapore, United States, Europe and Australia) and made its way to clinical adoption for ambulatory BP monitoring (ABPM; Table 1). However, as of 31 December 2025, HealthSTATS has wound up its operations, ceased support for BPro, and all regulatory approvals have since been withdrawn [2].

Table 1 Examples of pharmacies or clinics that offered BPro for ambulatory blood pressure monitoring servicesa.

Full size table

This is a preview of subscription content, access via your institution

Access options

Access through your institution

Subscribe to this journal

Receive 12 digital issues and online access to articles

111,21 € per year

only 9,27 € per issue

Buy this article

  • Purchase on SpringerLink
  • Instant access to the full article PDF.

39,95 €

Prices may be subject to local taxes which are calculated during checkout

Additional access options:

References

  1. Nair D, Tan SY, Gan HW, Lim SF, Tan J, Zhu M, et al. The use of ambulatory tonometric radial arterial wave capture to measure ambulatory blood pressure: the validation of a novel wrist-bound device in adults. J Hum Hypertens. 2008;22:220–2. https://doi.org/10.1038/sj.jhh.1002306

    Article  CAS  PubMed  Google Scholar 

  2. Healthstats BPro Evo [Internet]. [cited 2026 Jun 2]. Healthstats International | Ambulatory Blood Pressure Monitoring. Available from: https://www.healthstats.com

  3. IEEE Standard for Wearable, Cuffless Blood Pressure Measuring Devices. IEEE Std 1708-2025 (Revision of IEEE Std 1708-2014). 2026 Apr;1–72. https://doi.org/10.1109/IEEESTD.2026.11475616

  4. Stergiou GS, Avolio AP, Palatini P, Kyriakoulis KG, Schutte AE, Mieke S, et al. European Society of Hypertension recommendations for the validation of cuffless blood pressure measuring devices: European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability. J Hypertens. 2023;41:2074–87. https://doi.org/10.1097/HJH.0000000000003483

    Article  CAS  PubMed  Google Scholar 

  5. Williams B, Lacy PS, Baschiera F, Brunel P, Düsing R. Novel description of the 24-hour circadian rhythms of brachial versus central aortic blood pressure and the impact of blood pressure treatment in a randomized controlled clinical trial: The Ambulatory Central Aortic Pressure (AmCAP) Study. Hypertension. 2013;61:1168–76. https://doi.org/10.1161/HYPERTENSIONAHA.111.00763

    Article  CAS  PubMed  Google Scholar 

  6. Komori T, Eguchi K, Hoshide S, Williams B, Kario K. Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use. Blood Press Monit. 2013;18:57–62. https://doi.org/10.1097/MBP.0b013e32835d124f

    Article  PubMed  Google Scholar 

  7. Mukkamala R, Yavarimanesh M, Natarajan K, Hahn JO, Kyriakoulis KG, Avolio AP, et al. Evaluation of the accuracy of cuffless blood pressure measurement devices: challenges and proposals. Hypertension. 2021;78:1161–7. https://doi.org/10.1161/HYPERTENSIONAHA.121.17747

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. Kawaja A, Moosa AS, Lee EKP, Phoon IKY, Ang ATW, Chang ZY, et al. Comparison of patients’ acceptance of cuff-based vs wearable 24-hour ambulatory blood pressure monitoring devices: A multi-method study. PLoS One. 2025;20:e0336961 https://doi.org/10.1371/journal.pone.0336961

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Karow B. WHOOP, There It Is: lessons from WHOOP’s FDA warning letter. J Med Internet Res. 2026;28:e90882 https://doi.org/10.2196/90882

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Funding

The authors declare that no relevant funds, grants or other support were received during the preparation of this manuscript.

Author information

Authors and Affiliations

  1. The George Institute for Global Health, University of New South Wales, Sydney, NSW, Australia

    Isabella Tan & Sonali R. Gnanenthiran

  2. Macquarie Medical School, Faculty of Medicine, Health and Human Sciences, Macquarie University, Sydney, NSW, Australia

    Isabella Tan, Alberto P. Avolio & Mark Butlin

  3. Department of Cardiology, Concord Repatriation General Hospital, Sydney, NSW, Australia

    Sonali R. Gnanenthiran

  4. College of Medicine and Public Health, Flinders University, Adelaide, SA, Australia

    Genevieve M. Gabb

Authors

  1. Isabella Tan
  2. Alberto P. Avolio
  3. Mark Butlin
  4. Sonali R. Gnanenthiran
  5. Genevieve M. Gabb

Contributions

IT conceptualized, drafted and revised this manuscript. AA, MB, SG, and GG critically reviewed and revised this manuscript. All authors have read and approved the final manuscript.

Corresponding author

Correspondence to Isabella Tan.

Ethics declarations

Competing interests

AP Avolio is an advisor for CONNEQT Health Limited. The other authors have no competing interests to declare.

Additional information

Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Disclosures

AP Avolio is an advisor for CONNEQT Health Limited. The other authors have no conflicts of interests to declare.

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Tan, I., Avolio, A.P., Butlin, M. et al. A key player in the cuffless blood pressure devices market has exited – a cautionary tale for cuffless blood pressure monitoring?. J Hum Hypertens (2026). https://doi.org/10.1038/s41371-026-01240-8

Download citation

  • Received: 16 June 2026

  • Revised: 18 September 2026

  • Accepted: 23 September 2026

  • Published: 06 October 2026

  • Version of record: 06 October 2026

  • DOI: https://doi.org/10.1038/s41371-026-01240-8

来源:Nature 健康护理主题 · nature.com