多中心回顾性研究评估 AI 人工晶体计算公式与可解释机器学习用于白内障屈光预测
Multicenter evaluation of AI-based intraocular lens formulas and interpretable machine learning for cataract refractive prediction
一项多中心回顾性研究评估了 AI 人工晶体(IOL)计算公式与可解释机器学习模型对白内障术后屈光的预测表现,开发队列纳入四川大学华西医院 1203 只眼,外部验证队列纳入成都爱迪眼科医院 336 只高度近视眼。
- Article
- Open access
- Published: 22 October 2026
- Ning Yang1,2,
- Jiancheng Mu1,
- Yi Liu2,
- Wanyue Guo1,
- Chuhuan Sun1,
- Bosen Peng1,
- Junhan Xiang1,
- Lixing Deng1 &
- …
- Wei Fan1
Scientific Reports (2026) Cite this article
We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.
Abstract
Accurate intraocular lens (IOL) power selection remains challenging in cataract eyes with long axial length, altered corneal geometry, or previous corneal refractive surgery. This multicenter retrospective study evaluated conventional, post-refractive, and AI-based IOL formulas and developed interpretable machine learning models for postoperative refractive prediction. The development cohort included 1203 eyes from West China Hospital; the external validation cohort included 336 high-myopia eyes from Chengdu Aidi Eye Hospital, with 257 eyes available for cleaned counterfactual target-achievement analysis. Formula and model performance were assessed using mean absolute error, root mean square error, median absolute error, and predefined refractive error thresholds. In eyes without prior corneal refractive surgery, AI-based formulas outperformed conventional formulas, with Hill-RBF showing the lowest internal prediction error. Locally developed linear regression, support vector regression, and random forest models achieved comparable internal accuracy, but external validation showed reduced portability. SHAP analysis identified target refraction, axial length, and IOL power as dominant predictors. Post-refractive eyes, very long eyes, and flat corneas remained difficult. These findings support cautious, locally validated use of AI-based formulas and interpretable machine learning in refractive cataract surgery.
Abbreviations
- ACD:
-
Anterior chamber depth
- AL:
-
Axial length
- ATN:
-
Atrophy/traction/neovascularization
- BUII:
-
Barrett Universal II
- CCT:
-
Central corneal thickness
- IOL:
-
Intraocular lens
- LT:
-
Lens thickness
- MAE:
-
Mean absolute error
- MedAE:
-
Median absolute error
- MICE:
-
Multiple imputation by chained equations
- ML:
-
Machine learning
- RMSE:
-
Root mean square error
- SE:
-
Spherical equivalent
- SHAP:
-
SHapley Additive exPlanations
- SVR:
-
Support vector regression
- WTW:
-
White-to-white
Acknowledgements
The authors thank the clinical and technical staff of West China Hospital of Sichuan University and Chengdu Aidi Eye Hospital for their support in data collection and external validation.
Funding
This study received no specific grant from any funding agency in the public, commercial, or not-for-profit sectors.
Author information
Authors and Affiliations
Department of Ophthalmology, West China Hospital of Sichuan University, No. 37, Guoxue Alley, Wuhou District, Chengdu, 610041, Sichuan, China
Ning Yang, Jiancheng Mu, Wanyue Guo, Chuhuan Sun, Bosen Peng, Junhan Xiang, Lixing Deng & Wei Fan
Chengdu AIDI Eye Hospital, No. 41, Sec. 2, West 2nd Ring Road, Chengdu, 610072, Sichuan, China
Ning Yang & Yi Liu
Authors
- Ning Yang
- Jiancheng Mu
- Yi Liu
- Wanyue Guo
- Chuhuan Sun
- Bosen Peng
- Junhan Xiang
- Lixing Deng
- Wei Fan
Corresponding author
Correspondence to Wei Fan.
Ethics declarations
Ethics approval and consent to participate
The study was approved by the Medical Ethics Committee of West China Hospital, Sichuan University (approval number: 2024 (2514)). The study was conducted in accordance with the Declaration of Helsinki. Data were de-identified for analysis; consent waiver or additional consent was applied as permitted by the ethics committee.
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Yang, N., Mu, J., Liu, Y. et al. Multicenter evaluation of AI-based intraocular lens formulas and interpretable machine learning for cataract refractive prediction. Sci Rep (2026). https://doi.org/10.1038/s41598-026-76755-8
Received: 18 June 2026
Accepted: 14 October 2026
Published: 22 October 2026
DOI: https://doi.org/10.1038/s41598-026-76755-8
Keywords
来源:Nature 健康护理主题 · nature.com