跳到正文
原文
Medical Xpress·· 24 天前AI 评分45

大气固体分析探针质谱(ASAP-MS)现场快速筛查或有助于保护酒吧员工免受隐性毒品暴露

Rapid on-site screening could help protect bar staff from hidden drug exposure

AI 导读

英国萨里大学团队用 ASAP-MS 在一家营业中的英国酒吧开展现场筛查,3 周采集 420 份表面拭子,在卫生间、桌面和吧台检出可卡因与氯胺酮痕迹,单次分析不足 2 分钟。以实验室确证检测为对照,溶剂萃取的靶向方法对可卡因诊断准确率为 64.29%、氯胺酮为 83.81%,最低检出量分别为 20 pg/µL 和 0.78 ng/µL。

正文
Rapid on-site screening could help protect bar staff from hidden drug exposure
Heatmap of the number of positive detections for cocaine and ketamine in all four locations in sampling weeks 1–3 using a single-quadrupole LC–MS system. Credit: Journal of Mass Spectrometry (2026). DOI: 10.1002/jms.70121

The study, published in the Journal of Mass Spectrometry, tested the method in a working U.K. pub and found traces of cocaine and ketamine on surfaces across bathrooms, tables and bar areas.

The technique, atmospheric solids analysis probe mass spectrometry (ASAP-MS), skips the chromatographic separation step that traditional methods require—where a sample is first sorted into its individual ingredients before being identified in a lab. A sample is introduced via a glass capillary rod into a heated source, vaporized by nitrogen gas, and ionized without separation.

Analysis takes under two minutes, and that speed and simplicity could make rapid, on-site drug screening practical for venues, crime scenes or other locations where quick turnaround matters.

Andriana Michailidis, lead author and Ph.D. student at the University of Surrey, said, "Whether we like it or not, drugs are part of daily life and are often present where nobody can see them. Behind every one of these traces are people—staff who clean those bathrooms and work those shifts, and customers who have no idea what is on the surfaces they touch. When a venue can find out within minutes what is there, action can be taken that same night to protect them. Staff don't have to wait days for a laboratory result while the risk goes on unseen."

Over three weeks, the Surrey team collected 420 surface swabs from a single pub in the south of England. They tested three distinct workflows—direct untargeted analysis of swabs with no sample preparation, untargeted analysis of solvent-extracted samples, and targeted analysis monitoring specific chemical signatures. The targeted method using solvent-extracted samples achieved an overall diagnostic accuracy of 64.29% for cocaine and 83.81% for ketamine, compared against confirmatory laboratory testing.

The study found cocaine on all sampled bathroom surfaces using all three workflows. Concentrations were highest in bathroom areas, consistent with prior research showing that public bathrooms in venues are frequently used for substance consumption. The method identified very small amounts of cocaine and ketamine using the targeted approach.

The lowest levels were 20 picograms per microliter for cocaine and 0.78 nanograms per microliter for ketamine. Twenty picograms is about a millionth of the weight of a single human hair. Spread through a microliter, which is a droplet far smaller than a pinhead, it is vastly less than anyone could see or feel. The ketamine figure is about 40 times higher than the cocaine figure, but it is still roughly a billionth of a gram in a droplet too small to see clearly.

Dr. Patrick Sears, corresponding author from the University of Surrey's School of Chemistry and Chemical Engineering, said, "Atmospheric solids analysis probe mass spectrometry offers a practical bridge between the speed needed for field screening and the sensitivity required to detect drugs at trace levels. What excites us about this work is proving that a portable system can perform usefully in realistic, uncontrolled environments. This matters for venue safety, for occupational health monitoring of staff, and for understanding surface contamination where it actually occurs."

Ketamine proved more challenging to detect than cocaine because it remains trapped in cotton swab fibers and has lower volatility, meaning it vaporizes more slowly during analysis. Direct analysis of swabs missed significant amounts of ketamine that were later confirmed by traditional laboratory testing. Solvent extraction recovered those trapped compounds but added preparation time. The researchers identified 15 false positives across the three workflows, some attributed to procaine—a cutting agent occasionally mixed with cocaine—which produces a nearly identical chemical signature to ketamine.

The study is among the first to assess surface contamination in an operational venue and to test whether portable mass spectrometry could serve as a practical detection tool for public health or occupational safety applications.

More information

Andriana Michailidis et al., Assessing the use of ASAP‐MS for Screening a UK Pub for Illicit Drug Substances, Journal of Mass Spectrometry (2026). DOI: 10.1002/jms.70121

Key medical concepts

CocaineKetamine

Who's behind this story?

Gaby Clark

Gaby Clark

MA in English, copy editor since 2021 with experience in higher education and health content. Dedicated to trustworthy science news. Full profile →

Andrew Zinin

Andrew Zinin

Master's in physics with research experience. Long-time science news enthusiast. Plays key role in Science X's editorial success. Full profile →

Citation: Rapid on-site screening could help protect bar staff from hidden drug exposure (2026, October 27) retrieved 27 October 2026 from https://medicalxpress.com/news/2026-10-rapid-site-screening-bar-staff.html

This document is subject to copyright. Apart from any fair dealing for the purpose of private study or research, no part may be reproduced without the written permission. The content is provided for information purposes only.

来源:Medical Xpress · medicalxpress.com