社区环境下便携眼科诊断设备的实施与可行性研究:一项准实验前后研究
Community-based implementation and feasibility of portable ophthalmic diagnostic devices: a quasi-experimental pre-post study in underserved areas
该研究评估了两种便携式眼科诊断设备在欠发达农村地区的诊断一致性及实施可行性。在医院内,TopSlit 和 Reticapture 与传统设备的诊断一致性良好(κ 值为 0.535 至 0.944)。
- Article
- Open access
- Published: 03 November 2026
- Hoon Kim1,4,
- Jinho Kim2,4,
- Sun Kwon Kim5,
- Kyungsun Na3,
- Seok Bum Lee1,4,
- Won-Cheoul Jang2,4 &
- …
- Kyongjin Cho ORCID: orcid.org/0000-0003-1468-467X1,4
Scientific Reports (2026) Cite this article
We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.
Abstract
Portable ophthalmic diagnostic devices have emerged as promising tools to improve access to eye care in underserved communities; however, their agreement with conventional equipment and real-world implementation feasibility remain insufficiently evaluated. This study aimed to evaluate the diagnostic agreement of two portable ophthalmic devices and their real-world implementation feasibility in an underserved rural community. We conducted a two-phase study consisting of an in-hospital agreement study and a community-based implementation study. In the in-hospital phase, a smartphone-based portable slit lamp (TopSlit) and a handheld all-in-one imaging device (Reticapture) were compared with conventional stationary equipment in 72 patients with ocular diseases. Agreement was assessed using Cohen’s κ, intraclass correlation coefficients (ICC), and Spearman’s correlation. In the community-based phase, 124 residents aged ≥ 60 years from 10 underserved rural villages underwent baseline ophthalmic screening and a 6-month follow-up assessment. The primary endpoint was the 30-day hospital-visit completion rate among referred participants; secondary outcomes assessed implementation feasibility and participant acceptability, while longitudinal visual-acuity and patient-reported outcomes were exploratory. TopSlit and Reticapture showed moderate to almost perfect agreement with conventional reference equipment (κ = 0.535–0.944). Among 60 participants referred for further evaluation, the 30-day hospital-visit completion rate was 53.3% (32/60; 95% CI, 40.9–65.4%), exceeding the prespecified 40% feasibility threshold (exact p = 0.047). No upload/network or image-quality failures occurred during field implementation, and no additional image reacquisition or remote-grader requests for additional images were required. Participant satisfaction with the program was high (92.9%). These findings support the feasibility and acceptability of integrating portable ophthalmic devices into a community-based eye-care program in an underserved rural setting. The observed agreement with conventional reference equipment, operational performance, and linkage to follow-up care support further evaluation in larger, rigorously controlled studies. The study was approved by the Dankook University Hospital IRB (DKUH IRB 2022-10-016 (Date of Registration: 2022-10-13; Date of Approval: 2022-11-10) and DKUH IRB 2024-06-016 (Date of Registration: 2024-06-13; Date of Approval: 2024-08-02)). The study was registered at the Clinical Research Information Service (CRIS), Republic of Korea (https://cris.nih.go.kr; KCT0012177).
Funding
The present study was supported by the research fund of Dankook University in 2025.
Author information
Authors and Affiliations
Dankook University Hospital, Cheonan, Korea
Hoon Kim, Seok Bum Lee & Kyongjin Cho
Dankook University, Cheonan, Korea
Jinho Kim & Won-Cheoul Jang
The Catholic University of Korea, Seoul, Korea
Kyungsun Na
Dankook Institute on Aging, Cheonan, Korea
Hoon Kim, Jinho Kim, Seok Bum Lee, Won-Cheoul Jang & Kyongjin Cho
Korea Electrotechnology Research Institute, Changwon, Korea
Sun Kwon Kim
Authors
- Hoon Kim
- Jinho Kim
- Sun Kwon Kim
- Kyungsun Na
- Seok Bum Lee
- Won-Cheoul Jang
- Kyongjin Cho
Corresponding author
Correspondence to Kyongjin Cho.
Ethics declarations
Competing interests
The authors declare no competing interests.
Reporting guideline
This study is reported in accordance with the Standards for Reporting Implementation Studies (StaRI) statement and the Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) guidelines.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Kim, H., Kim, J., Kim, S.K. et al. Community-based implementation and feasibility of portable ophthalmic diagnostic devices: a quasi-experimental pre-post study in underserved areas. Sci Rep (2026). https://doi.org/10.1038/s41598-026-78353-0
Received: 13 July 2026
Accepted: 26 October 2026
Published: 03 November 2026
DOI: https://doi.org/10.1038/s41598-026-78353-0