基于物理信息的神经网络算子实现经颅磁刺激电场快速预测
Physics-informed neural operators for rapid electric field prediction in transcranial magnetic stimulation of healthy and tumor subjects
研究提出首个物理信息神经网络算子(PINO),用于个体化经颅磁刺激电场的快速预测。该模型在75名受试者上训练,可在消费级GPU不到100毫秒内预测约30万个灰质节点电场。在健康及肿瘤患者测试集中,其焦点内误差和均方根误差与近期深度学习结果相当,为实时电场引导治疗提供了技术基础。
- Article
- Open access
- Published: 16 October 2026
Scientific Reports (2026) Cite this article
We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.
Abstract
Accurate and rapid prediction of transcranial magnetic stimulation (TMS)-induced electric fields is essential for personalized clinical applications, from individualized therapy in major depressive disorder to presurgical functional mapping in neuro-oncology. Yet conventional finite element methods are often too computationally expensive for interactive use. Existing deep learning surrogate models partly address this limitation but are largely constrained to fixed voxel-grid representations that cannot naturally accommodate the irregular geometry of subject-specific head meshes. We present the first physics-informed neural operator (PINO) for subject-specific prediction of vector TMS electric fields, built on a mesh-native dual-encoder architecture that jointly encodes head anatomy and coil configuration. By incorporating a curl-based regularization term aligned with the curl-free nature of the TMS secondary field, the PINO model promotes physical consistency as a soft regularizer alongside data-driven supervision. Trained on 75 subjects from the IXI dataset, the framework was evaluated across four independent held-out test datasets, including healthy subjects from two additional MRI cohorts and, for the first time in learning-based TMS electric-field prediction, tumor patients. The resolution-agnostic formulation generalizes across subjects, coil types, and spatial resolutions without retraining, predicting vector electric fields on approximately 300,000 gray matter volume nodes in under 100 ms on a consumer-grade GPU. Across all test datasets within the focality-50% region, the PINO model achieved median center-of-gravity errors of 0.39–0.43 mm, RMSE of 0.12–0.16 V/m, and mean angular deviations of 6.7°–7.3°, with performance within or below the error ranges reported by recent deep-learning approaches. At the selected 5% physics-loss weighting, curl loss decreased by 9.1% relative to the data-only model, while RMSE and MAD increased by only 1.6% and 1.3%, respectively. These results suggest that physics-informed neural operators provide a scalable technical foundation for rapid gray-matter electric-field approximation, supporting future studies on real-time electric-field-guided TMS workflows
Acknowledgements
The authors acknowledge Claude (Anthropic) for assisting with language editing and improving clarity.
Funding
This work is financed by research subsidies granted by the government of Upper Austria (research projects MIMAS.ai, mimas.ng) and by the FFG (research project nARvibrain, grant no. 894756). www.ffg.at. “ICT of the Future” programme – an initiative of the Federal Ministry for Climate Action, Environment, Energy, Mobility, Innovation and Technology (BMK). RISC Software GmbH is a member of UAR (Upper Austrian Research) Innovation Network.
Author information
Authors and Affiliations
Research Unit Medical Informatics, RISC Software GmbH, Hagenberg, 4232, Austria
Philipp Moser, Fabian Greifeneder, Dominik Freinberger & Stefan Thumfart
Department of Radiology, Medical University of Graz, Graz, 8036, Austria
Gernot Reishofer
Department of Radiology, Clinical Division of Neuroradiology, Vascular and Interventional Radiology, Medical University of Graz, Graz, 8036, Austria
Pia Ritter
cortEXplore GmbH, Perg, 4320, Austria
Louis Frank
Department of Neurosurgery, Medical University of Graz, Graz, 8036, Austria
Sascha Freigang & Kariem Mahdy Ali
Authors
- Philipp Moser
- Fabian Greifeneder
- Dominik Freinberger
- Gernot Reishofer
- Pia Ritter
- Louis Frank
- Stefan Thumfart
- Sascha Freigang
- Kariem Mahdy Ali
Corresponding author
Correspondence to Philipp Moser.
Ethics declarations
Competing interests
The authors declare no competing interests.
Ethical approval
The use of data from both healthy subjects (MUG-TMS) and tumor patients (MUG-Tumor) collected at the Medical University of Graz was approved by the local ethics committee within the framework of the nARvibrain research project (approval number: ECS 115/2024). Informed consent was obtained from all subjects. The publicly available IXI Guy’s Hospital and Kirby21 datasets were used in accordance with their respective data usage terms. All procedures were conducted in accordance with the ethical principles of the Declaration of Helsinki.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Moser, P., Greifeneder, F., Freinberger, D. et al. Physics-informed neural operators for rapid electric field prediction in transcranial magnetic stimulation of healthy and tumor subjects. Sci Rep (2026). https://doi.org/10.1038/s41598-026-74936-z
Received: 19 June 2026
Accepted: 04 October 2026
Published: 16 October 2026
DOI: https://doi.org/10.1038/s41598-026-74936-z
Keywords
来源:Nature 健康护理主题 · nature.com