iFalcon V700 回弹式眼压计在离体兔眼中的多角度验证:与植入式压力传感器的对比研究

Multi-angle validation of the iFalcon V700 rebound tonometer against implantable pressure sensor in ex-vivo rabbit eyes

AI 导读

iFalcon V700 回弹式眼压计在 0–90° 空间倾斜范围内,与植入式压力传感器在 8–50 mmHg 眼压区间内显示出高度一致性,绝对中位差在 0.34–1.35 mmHg 之间,线性相关性强(R² = 0.962–0.989)。

正文

Scientific Reports (2026) Cite this article

We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.

Abstract

Rabbits are widely employed as ophthalmic models due to their favorable corneal geometry and biomechanical properties; however, intraocular pressure (IOP) measurements in both clinical and experimental settings are frequently performed under non-ideal body positions that may introduce off-axis spatial orientations relative to gravity. Despite this reality, no rebound tonometer has been validated for multi-angle use in rabbit eyes, leaving the impact of spatial inclination and gravitational deviation on measurement accuracy unknown. To validate the performance of the iFalcon V700 rebound tonometer across a 0–90° spatial inclination range (while maintaining strict perpendicularity between the probe and cornea) using an implanted pressure sensor as the gold standard. Experimental laboratory investigation. Four ex-vivo rabbit eyes. A motorized mechanical assembly facilitated multi-angle positioning of the eyes at seven distinct angles from 0° to 90° using 15° increments. An adjustable lead screw was used to gradually increase pressure to generate a physiologically relevant IOP range (8–50 mmHg). Simultaneous measurements were acquired using the iFalcon V700 tonometer (in rabbit-specific mode) and the implanted sensor, with precise synchronization ensured by a host computer program. Data distribution, linear regression, and Bland-Altman analyses were performed. IOP measurement agreement, correlation, and range-dependent biases between the rebound tonometer and the implanted sensor across various angles. Violin plot distributions of the iFalcon V700 closely mirrored the implanted sensor across all angles, with absolute median differences (|Δ Median|) remaining clinically negligible (range: 0.34–1.35 mmHg) and no significant measurement differences detected (P = 0.398–0.973). Regression analysis revealed a strong positive correlation (R² = 0.962–0.989). However, range-dependent biases were present; a negative intercept suggested underestimation at lower IOP, whereas a slope > 1 (range: 1.00–1.11) indicated systematic overestimation at higher pressures. Bland-Altman analysis corroborated this slight overestimation bias, demonstrating mean differences ranging from 0.05 mmHg (at 0°) to 1.12 mmHg (at 90°), with 95% limits of agreement tightly confined, reaching a maximum width of -4.46 to 5.89 mmHg at 45°. Validated against an implantable pressure sensor, the iFalcon V700 rebound tonometer demonstrated high agreement across the 8–50 mmHg range and maintained robust mechanical and algorithmic stability throughout 0–90° multi-angle measurements under controlled ex vivo conditions. Despite range-dependent overestimation at higher pressures, this study, heavily constrained by its small sample size (n = 4 eyes), must be interpreted strictly as a preliminary pilot study and an engineering proof-of-concept for multi-angle tonometry. Therefore, adequately powered, large-scale validation in living animals and clinical populations is definitively required before any translational claims regarding dynamic positional monitoring can be established.

Acknowledgements

Not applicable.

Funding

This work was supported by the Zhejiang Province Leading Geese Plan (Grant No. 2026C02A1189) and the Science Foundation of Zhejiang Sci-Tech University (Grant No. 24242151-Y).

Author information

Authors and Affiliations

  1. School of Mechanical Engineering, Zhejiang Sci-Tech University, Hangzhou, 310018, China

    Xinqi Cui, Yingbo Liu, Qianru Chen & Ji Fu

  2. Rui Pai Hong Tai Pet Hospital, Huancheng Main Hospital, Hangzhou, 310000, China

    Lihui Yang

Authors

  1. Xinqi Cui
  2. Yingbo Liu
  3. Lihui Yang
  4. Qianru Chen
  5. Ji Fu

Corresponding author

Correspondence to Ji Fu.

Ethics declarations

Competing interests

The authors declare no competing interests.

Consent for publication

Not applicable.

Protocol registration

Not applicable.

Compliance and ethics

The authors declare that they have no conflict of interest. The animal experiments and animal handling procedures were approved by the Institutional Animal Care and Use Committee of Hangzhou Lifu-Tai biotechnology co., Ltd (number of permit: IACUC-20260417-01). The experiments followed the relevant guidelines of the National Institutes of Health (NIH) and strictly followed internationally recognized animal welfare and ethical guidelines, aiming to manage and use experimental animals in a humane and scientific manner. The animals used in the experiments (New Zealand white rabbits) were provided by Hangzhou Lifu-Tai biotechnology co., Ltd (experimental animal license number: SCXK (Zhe) 2023-0001). All rabbits were confirmed free of ocular disease (e.g., corneal lesions, lens opacities, glaucoma) by slit-lamp examination and indirect ophthalmoscopy. Following euthanasia (performed in accordance with the AVMA *Guidelines for the Euthanasia of Animals*), the eyeballs were enucleated, and IOP measurements were taken ex-vivo to avoid interference caused by rabbit agitation and to minimize stress during the measurement process. The experiment adhered to the guidelines of the Association for Assessment and Accreditation of Laboratory Animal Care (AAALAC) and the ARRIVE guidelines. Euthanasia was induced via intravenous injection of sodium pentobarbital (100 mg/kg) using a medical syringe, and IOP measurements began within 5 min post-mortem.

Declaration of generative AI use

None of the content is directly generated by AI.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Cui, X., Liu, Y., Yang, L. et al. Multi-angle validation of the iFalcon V700 rebound tonometer against implantable pressure sensor in ex-vivo rabbit eyes. Sci Rep (2026). https://doi.org/10.1038/s41598-026-76728-x

Download citation

  • Received: 05 June 2026

  • Accepted: 14 October 2026

  • Published: 21 October 2026

  • DOI: https://doi.org/10.1038/s41598-026-76728-x

Keywords

来源:Nature 健康护理主题 · nature.com