跳到正文
原文
Nature 健康护理主题·· 23 天前AI 评分32

基于快速非绝热磁场关断的集成 NMR 磁场探针用于超低场 MRI

An integrated NMR field probe using rapid non-adiabatic field switch-off for (ultra) low field MRI

AI 导读

研究团队发布一种用于超低场 MRI 的集成 NMR 磁场探针,利用高压碳化硅晶体管实现快速非绝热磁场关断。该探针无需预先知晓磁场强度即可测量拉莫尔频率,验证范围从 1 mT 到 45 mT,比类似系统强 1000 倍。其低成本模块化设计有助于解决磁场不稳定性问题,支持全球范围内低成本 MRI 的可复制开发与图像质量改进。

正文

Communications Engineering (2026) Cite this article

We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.

Abstract

Access to magnetic resonance imaging (MRI) remains severely limited in low- and middle-income countries, especially in sub-Saharan Africa, despite rising rates of non-communicable diseases. Low-field MRI presents an affordable, locally developable diagnostic solution, but its performance is constrained by magnetic field instability. We present a novel NMR field probe designed to overcome these challenges using a rapid non-adiabatic switch-off of a pre-polarization field resulting in precessing spin magnetization. Achieved by first use of high-voltage silicon carbide transistors operating in controlled avalanche breakdown, it measures the Larmor frequency without prior field knowledge, unlike conventional probes. This capability is crucial during magnet development with often unknown fields, allowing early detection of magnet issues, and offering an urgently needed tool for magnet design and image-quality improvement. Validated from 1 mT to 45 mT, up to 1,000 times stronger than similar systems, its low-cost, modular design supports replication, upgrades, and enhanced field control, helping expand global MRI access.

Acknowledgements

P.P. would like to thank S. Eckardt (GreenTeam, University of Stuttgart) for his help in embedded coding. Support by the Max Planck Society is greatly acknowledged.

Funding

Funding by the European Research Council (ERC Advanced Grant No 834940, SpreadMRI) is gratefully acknowledged. K.B. acknowledges funding from the DFG (BU 2694/6-1, BU 2694/9-1, 469366436, 527345502). Open Access funding enabled and organized by Projekt DEAL.

Author information

Author notes

  1. These authors contributed equally: Pavel Povolni, Dominique Goerner.

Authors and Affiliations

  1. High‑Field Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany

    Pavel Povolni, Dominique Goerner, Praveen Iyyappan Valsala, Georgiy Solomakha, Nicolas Kempf, Felix Glang, Judith Samlow, Kai Buckenmaier & Klaus Scheffler

  2. Institute of Robust Power Semiconductor Systems ILH, University of Stuttgart, Stuttgart, Germany

    Dominique Goerner, Lukas Gebert, Ruben Schnitzler & Ingmar Kallfass

  3. Department for Biomedical Magnetic Resonance, University of Tübingen, Tübingen, Germany

    Praveen Iyyappan Valsala & Klaus Scheffler

  4. Institute of Biomedical Imaging, Graz University of Technology, Graz, Austria

    Felix Glang

  5. Hospital Diospi Suyana, Curahuasi, Apurímac, Peru

    Judith Samlow

Authors

  1. Pavel Povolni
  2. Dominique Goerner
  3. Praveen Iyyappan Valsala
  4. Lukas Gebert
  5. Georgiy Solomakha
  6. Nicolas Kempf
  7. Felix Glang
  8. Judith Samlow
  9. Ruben Schnitzler
  10. Ingmar Kallfass
  11. Kai Buckenmaier
  12. Klaus Scheffler

Corresponding authors

Correspondence to Pavel Povolni or Kai Buckenmaier.

Ethics declarations

Competing interests

D.G. works as an engineer for Swabian Instruments (Stuttgart, Germany, development of digital data acquisition hardware) in the meantime. Most of this work was done before his employment, as well as in his spare time. The research was completely funded by the MPI and other grants awarded. All other authors declare no conflict of interest.

Additional information

Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Povolni, P., Goerner, D., Iyyappan Valsala, P. et al. An integrated NMR field probe using rapid non-adiabatic field switch-off for (ultra) low field MRI. Commun Eng (2026). https://doi.org/10.1038/s44172-026-00815-0

Download citation

  • Received: 18 December 2025

  • Accepted: 09 October 2026

  • Published: 28 October 2026

  • DOI: https://doi.org/10.1038/s44172-026-00815-0