一项前瞻性随机对照试验纳入126例接受下颌前磨牙根管再治疗的患者,比较传统电子根尖定位仪(Root ZX Mini)与三种集成工作长度控制模式(apical reverse、apical slow down、apical stop)的术后疼痛。

正文

Scientific Reports (2026) Cite this article

We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.

Abstract

The aim of this study was to evaluate the effect of different simultaneous working length control modes of an integrated electronic apex locator (EAL) system on postoperative pain following retreatment compared with working length determination using a conventional EAL. This prospective, parallel-group, randomized controlled clinical trial included 144 patients undergoing nonsurgical retreatment of mandibular premolars, of whom 126 were included in the final analysis. Patients were randomly allocated to 4 groups: conventional EAL (control; Root ZX Mini) and three integrated EAL modes (apical reverse, apical slow down, and apical stop). Postoperative pain intensity was recorded at seven predefined time points: 6 and 12 h and days 1, 2, 3, 5, and 7 after treatment, using the Numeric Rating Scale (NRS). A significant main effect of group was observed (p = 0.006). Averaged across assessment points, NRS scores were higher in the conventional EAL group than in the apical reverse group (p = 0.010). Other pairwise group differences were not statistically significant (p > 0.05). A significant main effect of time was also observed (p < 0.001), whereas the group × time interaction was not significant (p = 0.077). Procedure time was significantly longer in the conventional EAL group (p < 0.001). In univariate analyses, longer procedure time was associated with higher odds of postoperative pain at all evaluated time points (p < 0.05). The apical reverse mode was associated with lower overall postoperative NRS scores than conventional working length determination, while all three integrated EAL modes were associated with shorter procedure times. No statistically significant difference in pain trajectories was detected among the groups.

Acknowledgements

We thank all study participants. This article was derived from Hasan Uyanık’s dissertation.

Funding

This study has been supported by the Recep Tayyip Erdoğan University Development Foundation (Grant no. 020260070270534).

Author information

Authors and Affiliations

  1. Department of Endodontics, Faculty of Dentistry, Recep Tayyip Erdoğan University, Rize, Turkey

    Hasan Uyanık & Edanur Maraş

Authors

  1. Hasan Uyanık
  2. Edanur Maraş

Corresponding author

Correspondence to Edanur Maraş.

Ethics declarations

Competing interests

The authors declare no competing interests.

Ethics approval

The study was approved by the local Ethics Committee (Approval No: 2025/113) and the Republic of Türkiye Ministry of Health, Turkish Medicines and Medical Devices Agency (Approval No: E-24931227-514.13.02-5708670). The study was retrospectively registered at ClinicalTrials.gov (NCT07527897; first registered on 14 April 2026).

Consent to participate

Informed consent was obtained from all participants included in the study.

Additional information

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Uyanık, H., Maraş, E. Postoperative pain after endodontic retreatment using different integrated electronic apex locator working length control modes. Sci Rep (2026). https://doi.org/10.1038/s41598-026-80660-5

Download citation

  • Received: 01 July 2026

  • Accepted: 10 November 2026

  • Published: 14 November 2026

  • DOI: https://doi.org/10.1038/s41598-026-80660-5

Keywords

来源:Nature 健康护理主题 · nature.com