交联透明质酸真皮填充剂临床前评价:从内在特性到情境依赖性表现的关键考量

Key considerations from intrinsic properties to context-dependent performance in the preclinical evaluation of crosslinked HA dermal fillers

AI 导读

一项发表于 Scientific Reports 的临床前研究对5种交联透明质酸真皮填充剂(含 Juvéderm VOLUMA、Restylane Defyne、Ifresh 及两种 SPDHA 凝胶)进行多指标表征,发现内在网络参数无法单独预测情境依赖性表现。Restylane Defyne 的 γc 最高但分散时间最短,作者指出体外相关性不能替代临床证据,仍需体内和临床研究验证。

正文

Scientific Reports (2026) Cite this article

We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.

Abstract

Current preclinical evaluation of crosslinked hyaluronic acid (HA) dermal fillers often treats individual assays—rheology, extrusion force, cohesivity, swelling—in isolation, making it difficult to relate intrinsic network properties to context-dependent performance. In this study, five crosslinked HA gels—three commercial BDDE-crosslinked fillers (Juvéderm VOLUMA, Restylane Defyne and Ifresh) and two laboratory-model spermidine-crosslinked HA (SPDHA) gels—were characterized using a multi-assay battery covering extrusion force measurement, rheological characterization (amplitude sweeps, frequency/time sweeps, thixotropy test and flow sweep), cohesivity assessment (drop weight test and dispersion time test), and water uptake evaluation (swelling factor and hydration test). We examined how these different metrics relate to each other. Our results reveal that intrinsic network properties—such as elastic modulus (G′), crossover strain (γc) and swelling factor—capture different aspects of gel behavior from context-dependent performance metrics—such as extrusion force, dispersion time (reflecting resistance to hydrodynamic shear), and confined hydration (simulating post-injection spatial constraint). For example, Restylane Defyne exhibited the highest γc among all products, indicating superior intrinsic resistance to large deformation, yet it showed the shortest dispersion time—a context-dependent outcome. Such discrepancies demonstrate that no single intrinsic parameter can reliably predict context-dependent performance. Interpreting the two sets of metrics in conjunction provides a more comprehensive understanding of each gel’s behavior, which may assist clinicians in product selection and guide future clinical investigation. However, correlations obtained in vitro cannot substitute clinical evidence; further in vivo and clinical studies are necessary to validate these relationships.

Acknowledgements

The authors thank their colleagues for technical assistance and insightful discussions.

Funding

Authors acknowledge funding by IMEIK Technology Development Co. Ltd.

Author information

Authors and Affiliations

  1. Beijing Engineering Lab of Neo-Biodegradable Materials, Beijing, 100022, China

    Ran Yu, Jiahua Cao & Kun Zhang

Authors

  1. Ran Yu
  2. Jiahua Cao
  3. Kun Zhang

Corresponding author

Correspondence to Kun Zhang.

Ethics declarations

Competing interests

All authors are researchers at the Beijing Engineering Lab of Neo-Biodegradable Materials and R&D engineers at IMEIK Technology Development Co., Ltd. Nonetheless, this affiliation has not influenced the scientific rigour or objectivity of the research presented in this manuscript. Unless stated otherwise, the soft-tissue fillers discussed in this manuscript were purchased from commercial sources. The chemical reagents used in this manuscript are standard laboratory reagents and do not carry any commercial endorsements.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Yu, R., Cao, J. & Zhang, K. Key considerations from intrinsic properties to context-dependent performance in the preclinical evaluation of crosslinked HA dermal fillers. Sci Rep (2026). https://doi.org/10.1038/s41598-026-77274-2

Download citation

  • Received: 11 May 2026

  • Accepted: 19 October 2026

  • Published: 28 October 2026

  • DOI: https://doi.org/10.1038/s41598-026-77274-2

Keywords

来源:Nature 健康护理主题 · nature.com