Effect of vancomycin-containing lubricant on vitamin E-stabilized UHMWPE wear in an in vitro total knee simulator: a pilot study
一项在ISO 14243-1:2009模拟器条件进行的初步研究显示,8个全膝置换组件分别置于含血清或含万古霉素血清润滑剂中。该研究旨在比较两种润滑条件下维生素E稳定UHMWPE衬垫的磨损率。
Introduction
The incidence of primary total knee arthroplasty (TKA) is increasing as a result of longer life expectancy and demographic shifts, which has been accompanied by a rise in procedure-related complications1. Periprosthetic joint infection (PJI) remains one of the most devastating complications, associated with significant morbidity and mortality, a substantial economic burden on healthcare systems worldwide as well as being the leading cause of early failure after knee replacement1,2,3,4.
Gram-positive bacteria are the predominant pathogens in PJI, with Staphylococcus epidermidis and Staphylococcus aureus as the most common. Additional risk factors include higher BMI, longer surgical time, longer hospitalization, immunosuppressants, preoperative hypoproteinemia, transfusion and superficial infection5. This highlights the need for continued efforts to reduce complication rates, ease healthcare system burden, and improve patient outcomes.
Systemic perioperative antibiotic prophylaxis remains the standard evidence-based approach to infection prevention in primary arthroplasty, with contemporary recommendations emphasizing appropriate pre-incision administration and limited duration6. In recent years, intraoperative topical vancomycin is widely used, whereas the biomechanical effect of topical vancomycin on the wear behavior remains unknown. By contrast, the addition of topical or intra-articular vancomycin in primary total knee arthroplasty remains investigational and is not established as routine care. Recent randomized evidence has not shown a significant reduction in periprosthetic joint infection or surgical-site infection and has raised concern regarding aseptic wound complications7,8. Vancomycin is a glycopeptide antibiotic with activity against many Gram-positive organisms9, but its local use in arthroplasty should therefore be described as an experimental adjunct rather than standard prophylaxis.
Beyond infection control, the long-term success of total joint arthroplasty depends on minimizing mechanical wear. Wear particles may induce inflammatory responses, osteolysis, aseptic loosening, and implant failure10. Although crosslinking and antioxidant stabilization have improved polyethylene performance, wear remains relevant to long-term implant durability11. Furthermore, the physical form and surrounding medium may influence tribological behavior, as dry powder applied to a surgical wound, an aqueous vancomycin solution, and vancomycin dissolved in protein-containing serum are not equivalent exposures. Vancomycin may alter protein-layer formation, viscosity, surface deposition, or other lubricant properties in addition to any direct interaction with UHMWPE.
The present experiment evaluated the combined effect of a vancomycin-containing serum lubricant versus control under force-controlled ISO 14243-1:2009 simulator conditions. This pilot study was positioned as an incremental addition to existing experimental evidence and was not prospectively designed as a non-inferiority or equivalence study. The exploratory objective was to describe and compare measured wear rates between the two lubricant conditions while explicitly considering protocol differences and the limited number of specimens.
Materials and methods
Prosthesis and embedment
A total of eight complete bicondylar total knee arthroplasty systems (5 C® Knee System, implantcast GmbH, Buxtehude, Germany) were used: four systems for the serum control condition and four for the vancomycin-containing lubricant condition. In each condition, three inserts were dynamically loaded and one insert served as the load-soak control; no additional unused systems were included in the reported experiment. All implants were size 7 L. The femoral components (5 C® MEDic Knee System) and tibial baseplates (5 C® Knee System) were manufactured from cobalt-chromium-molybdenum alloy in accordance with ISO 5832-4. The corresponding 7/9-mm tibial inserts (5 C® MEDic Knee System) were made from vitamin E-stabilized (implacross® E) UHMWPE (GUR; 1020; γ-sterilized ≥ 25 kGy and packed under inert gas. Manufacturing LOT numbers of components are provided in Supplementary Tables S1 and S2.
To obtain consistent alignment, the tibial components were embedded using custom-designed 3D-printed positioning guides. The guides produced a defined translational shift in the frontal plane; the extent of the shift was calculated from ISO 14243-1:200912. This embedding approach targeted a load distribution of 60% medial and 40% lateral, consistent with the mediolateral force ratios reported by Halder et al.13. Femoral components were embedded in polyurethane resin (RenCast® FC 52/53; Huntsman Advanced Materials, Basel, Switzerland) and mounted through a stainless-steel tube. Tibial components were embedded in the same resin and fixed with the standardized simulator fixture. Each chamber was enclosed with an elastic sealing membrane.
Wear simulator testing
Wear testing was performed using a customized four-station servo-hydraulic knee simulator (EndoLab GmbH, Thansau, Germany) under force-controlled ISO 14243-1:2009 conditions12 at a frequency of 1.0 Hz. The two lubricant conditions were tested in separate condition-specific runs because the simulator contained four stations. In each run, station 0 contained the load-soak control, while stations 1–3 contained the three dynamically loaded inserts (Fig. 1).
To minimize potential station-specific effects, the dynamically loaded inserts were rotated among stations 1–3 (Fig. 2). Station assignment was therefore not randomized but followed the predefined rotation sequence. No investigator blinding was applied during testing, cleaning, weighing, or visual assessment. The three dynamically loaded inserts within each lubricant condition represent technical replicates from a single simulator run rather than independent simulator runs.
Schematic illustration of the two condition-specific four-station simulator runs. In each run, three inserts (1–3) were dynamically loaded and one insert (0) served as the load-soak control.
Schematic representation of specimen rotation among stations 1–3 of the serum control group and vancomycin-containing lubricant group. Insert 0 is the load-soak control; inserts 1–3 are dynamically loaded and rotated.
The dynamic stations received the same axial load, flexion-extension, anteroposterior force, and internal-external torque profiles for 5.0 million cycles, whereas the load-soak station received axial loading without flexion, anteroposterior translation, or tibial torque (Fig. 3).
Load and kinematic patterns applied by the servo-hydraulic knee simulator in accordance with ISO 14243-1:2009, illustrating (A) axial load and flexion-angle trajectories and (B) tibial torque and anteroposterior force over one cycle. The curves were redrawn schematically from the standard rather than reproduced directly.
The base lubricant consisted of newborn calf serum (Biochrom GmbH, Berlin, Germany) diluted with distilled water to a protein concentration of 20 g/L. Each chamber contained 300 mL of lubricant. EDTA (24 g/L; AppliChem GmbH, Darmstadt, Germany) and amphotericin B (10 mL/L; Biochrom GmbH, Berlin, Germany) were added to limit calcium precipitation and fungal growth. In the exposed condition, 1 g of commercially supplied crystalline vancomycin powder (Eberth Arzneimittel GmbH, Ursensollen, Germany) was dissolved in 10 mL of distilled water and added to 290 mL of base lubricant, resulting in a final chamber volume of 300 mL and a nominal vancomycin concentration of 3.33 mg/mL. This approach was chosen to ensure complete dissolution of the vancomycin and to avoid potential third-body wear caused by undissolved particles. The complete lubricant, including vancomycin, was renewed every 0.5 million cycles. This prolonged repeated exposure represents an experimental upper-bound condition and does not reproduce a single intraoperative powder application. Vancomycin stability, pH, viscosity, osmolality, precipitation, and protein-layer formation were not measured and are acknowledged as uncharacterized exposure variables.
Before testing, all UHMWPE inserts were preconditioned in the base serum lubricant without vancomycin. Mass was monitored using an analytical balance (BP211D; Sartorius AG, Göttingen, Germany; readability ± 0.01 mg) in accordance with ISO 14243-2:2016. Conditioning was considered complete when the 24-hour mass change was less than 10% of the previous cumulative increase. Because inserts assigned to the exposed condition were not equilibrated in vancomycin-containing lubricant, preconditioning and test media were unmatched. Ongoing mass gain during the first 0.5 million cycles may therefore have influenced early corrected wear estimates.
For each dynamically loaded insert i at time t, gravimetric wear was calculated from buoyancy-corrected masses as Wi(t) = [mi(0) − mi(t)] + [mLS(t) − mLS(0)], where mLS denotes the condition-specific load-soak control. One load-soak trajectory was shared by all three dynamically loaded inserts in each run. Consequently, corrected observations within a condition contain correlated uncertainty. The original analysis did not propagate load-soak measurement uncertainty into group estimates; the revised interpretation therefore treats the specimen-level results as descriptive.
Gravimetric and surface assessment
The lubricant and all additives were renewed after every 0.5 million cycles. Inserts were removed for cleaning and gravimetric analysis at 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 million cycles. At each measurement interval, the dynamically loaded inserts and the load-soak control were cleaned, dried, and weighed in accordance with ISO 14243-2:2016. Briefly, the inserts were ultrasonically cleaned for 10 min in deionized water, rinsed with deionized water, ultrasonically cleaned for 10 min in a detergent solution prepared at the manufacturer-recommended concentration, and rinsed again. This was followed by ultrasonic cleaning in deionized water for 10 min and then for an additional 3 min, with a deionized-water rinse after each step. The inserts were subsequently dried, immersed in propan-2-ol for 5 min ± 15 s, dried using a filtered inert-gas jet, and placed under a vacuum of at least 13.3 Pa for a minimum of 30 min. Each insert was weighed five times within 90 min of removal from the vacuum chamber. At each of the five measurement time points, the weighing order of the inserts was randomized to minimize potential order-related measurement bias. If any two measurements differed by more than 0.1 mg, additional measurements were obtained until at least two readings agreed within 0.1 mg. Inserts were stored in sealed, dust-free containers between measurements.
In addition to gravimetric analysis, all three inserts in the vancomycin condition underwent simultaneous serial micro-CT examination after the 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 million cycle measurements using a micro-computed tomography system (micro-CT) (ProCon X-Ray GmbH, version 1.6.3.0; loaded device assembly: XTomDevice, version 1.7.0.0; Sarstedt, Germany. Scanning was performed at a tube voltage of 100 kV and an emission current of 210 µA, with an exposure time of 0.350 s, 2 × 2 detector binning, 200 averages, and 2,000 angular projections. Each scan required approximately 18 min 20 s, during which all inserts were scanned simultaneously. Solely the vancomycin-condition inserts underwent serial micro-CT examinations, which introduced additional imaging and out-of-fluid handling in that condition.
At the gravimetric assessment intervals, inserts were scanned using a flatbed scanner (Expression 1680 Pro; Seiko Epson Corporation, Tokyo, Japan) at 300 dpi with a 1-mm scale. Femoral and tibial components were photographed before and after simulation using a D5500 digital camera and DX 18–105-mm lens (Nikon Corporation, Tokyo, Japan) with a + 5-diopter macro converter. Surface images were reviewed qualitatively without a prespecified scoring system, independent blinded observers, profilometry, microscopy, or digital area quantification. Accordingly, the images are presented as descriptive illustrations and were not used for inferential between-group comparisons.
Data analysis and statistics
Between-condition wear-rate comparisons were performed descriptively because each lubricant condition was represented by a single condition-specific simulator run, with the three dynamically loaded inserts within each run treated as technical replicates rather than independent experimental units. Wear-rate data are therefore reported at the specimen level together with condition-specific mean ± standard deviation (SD) and the mean between-condition difference. No inferential statistical test was performed for the between-condition comparison, and no confidence intervals were calculated for the difference between conditions. No non-inferiority margin was prespecified, and no claims of non-inferiority or equivalence were made.
Cumulative gravimetric wear was characterized using specimen-level linear regression in GraphPad Prism 5 (GraphPad Software, San Diego, CA, USA). Regression analyses included measurements obtained at 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 million cycles, with intercepts left unconstrained. Regression slopes, representing estimated wear rates, and coefficients of determination (R2) were reported descriptively. No prespecified R2 threshold was used to define linearity.
Results
Wear behavior
The mean measured wear rate was 2.21 ± 0.92 mg/million cycles for the vancomycin-containing lubricant condition and 6.80 ± 3.07 mg/million cycles for the serum control condition, corresponding to a mean between-condition difference of − 4.59 mg/million cycles (vancomycin-containing lubricant minus serum control). All three serum control wear-rate estimates exceeded all three estimates obtained under vancomycin-containing lubricant.
Gravimetric wear data for both conditions before and after correction for the load-soak effect are shown in Figs. 4 and 5, respectively. Specimen-level linear regression lines and their corresponding 95% confidence intervals are presented in both figures, and regression slopes and R² values are summarized in Table 1.
Gravimetric wear of the UHMWPE inserts of the (A) serum control group and (B) vancomycin-containing lubricant before correction for the load-soak effect, including linear regression with 95% confidence intervals (dotted lines) and slope (m). A0 and D0 are the load-soak controls; A1–A3 are dynamically loaded serum control inserts; D1–D3 are dynamically loaded inserts exposed to vancomycin-containing lubricant.
Gravimetric wear of the UHMWPE inserts of the (A) serum control group and (B) vancomycin-containing lubricant after correction for the load-soak effect, including linear regression with 95% confidence intervals (dotted lines) and slope (m). A1–A3 are dynamically loaded serum control inserts; D1–D3 are dynamically loaded inserts exposed to vancomycin-containing lubricant.
Visual assessment of wear patterns
Insert wear
After 5.0 million cycles, all inserts showed visible surface changes, including burnishing and occasional scratches, without visible delamination or pitting (Fig. 6). The assessment was qualitative and unblinded; therefore, no quantitative claim regarding larger worn areas or a between-condition difference is made. A0 and D0 denote the load-soak controls and were not included in the dynamically loaded wear-rate comparison. A1–A3 denote the dynamically loaded serum control inserts, and D1–D3 denote the dynamically loaded inserts exposed to vancomycin-containing lubricant.
Qualitative surface images after 5.0 million cycles. A0 and D0 are the load-soak controls; A1–A3 are dynamically loaded serum control inserts; D1–D3 are dynamically loaded inserts exposed to vancomycin-containing lubricant. Encircled areas were identified visually and were not quantified.
Metal components
Femoral and tibial components in both conditions showed small scratches within the contact regions after 5.0 million cycles (Fig. 7). Because the assessment was qualitative, unblinded, and not based on predefined scoring or surface-roughness measurements, the images are reported descriptively without a formal between-condition comparison.
Qualitative images of femoral and tibial metal components after 5.0 million cycles. A0 and D0 correspond to the load-soak stations; specimens 1–3 were dynamically loaded. No quantitative surface-roughness or scratch-frequency analysis was performed.
Discussion
Under the tested conditions, the three inserts exposed to vancomycin-containing lubricant exhibited lower gravimetric wear estimates than the three inserts tested with serum control lubricant. However, the two conditions were evaluated in separate simulator runs and differed in preconditioning medium and micro-CT handling. In addition, each condition relied on a single shared load-soak control. Consequently, the observed between-condition difference may not be attributed specifically to vancomycin and should be interpreted as a descriptive finding.
This pilot study provides an incremental comparison rather than establishing a previously unexplored clinical concept. Published knee-simulator studies under related, but not identical, conditions report wear rates of a similar order of magnitude. Schroeder et al. reported 3.8 ± 0.5 mg/million cycles for antioxidant-blended crosslinked polyethylene under force-controlled ISO 14243-1 conditions14. Dreyer et al. reported approximately 1.5 mg/million cycles under force-controlled testing15, and Micheli et al. reported 2.4 ± 0.5 mg/million cycles for vitamin E-stabilized crosslinked UHMWPE under displacement-controlled conditions with antibiotic additives in the fluid16. These comparisons provide context but cannot resolve the protocol confounding in the present experiment.
Wear was more prominent on the medial compartment in both conditions. This pattern is consistent with asymmetric tibiofemoral loading, where approximately 60% of the load is applied medially and 40% laterally under typical gait13.
A mechanistic explanation for the lower measured wear estimates is not established. It is conceivable that vancomycin or vancomycin-related changes in the serum lubricant could alter adsorption, protein-layer formation, viscosity, surface deposition, or boundary lubrication17. Any proposed hydrophilic interaction between vancomycin and the hydrophobic UHMWPE matrix remains a hypothesis rather than a demonstrated mechanism. Targeted tribochemical studies with fluid characterization and quantitative surface analysis are required.
The group-specific micro-CT protocol is a major source of potential confounding. Only the vancomycin-condition inserts underwent repeated scanning and additional out-of-fluid handling. Even if the radiation dose was far below that used for intentional UHMWPE crosslinking18, the combined effects of imaging, oxygen exposure, handling, drying, and delayed return to the lubricant cannot be separated from the lubricant condition.
No post hoc power calculation is presented. Observed power is a function of the observed p-value and does not provide useful additional information19. A future confirmatory experiment should use an external variance estimate, a prespecified clinically or tribologically meaningful difference, replicated and counterbalanced simulator runs, and a prospectively defined non-inferiority margin.
The present wear experiment cannot establish infection-prevention efficacy, overall clinical safety, or support routine clinical use of topical vancomycin. Furthermore, the available clinical evidence remains inconclusive20,21. A 2024 double-blind randomized trial in primary total knee arthroplasty found no reduction in periprosthetic joint infection or surgical-site infection and reported more aseptic wound complications with intrawound vancomycin7. A 2026 multicenter randomized trial likewise found no significant one-year difference in infection outcomes across vancomycin powder, dilute povidone-iodine, combined, and saline-control strategies in high-risk primary arthroplasty8. The present study should therefore be interpreted only as a preliminary tribological observation under a prolonged vancomycin-containing lubricant exposure.
Future work should use protocols that differ only in vancomycin exposure, with matched preconditioning and test media, replicated and counterbalanced simulator runs, prespecified blinding, complete calibration documeentation, and independent load-soak controls or statistical propagation of shared-control uncertainty. Quantitative surface methods such as blinded scoring, digital area measurement, profilometry, microscopy, or roughness analysis should replace subjective visual comparisons. Exposure characterization should include concentration, stability, pH, viscosity, osmolality, temperature, precipitation, and protein-layer changes.
Conclusion
Three dynamically loaded inserts exposed to vancomycin-containing lubricant showed lower measured wear than three serum control inserts. However, the small sample, separate condition-specific runs, technical-replicate structure, unmatched preconditioning, shared load-soak correction, and group-specific micro-CT protocol prevent attribution of this difference to vancomycin. Therefore, the findings are descriptive and do not establish non-inferiority, clinical safety, infection-prevention efficacy, or support for routine topical use.
Data availability
The complete uncorrected and corrected mass measurements, load-soak trajectories, regression outputs, surface images, and analysis code are with the journal.
References
Kamath, A. F. et al. Quantifying the burden of revision total joint arthroplasty for periprosthetic infection. J. Arthroplasty. 30 (9), 1492–1497. https://doi.org/10.1016/j.arth.2015.03.035 (2015).
Premkumar, A. et al. Projected economic burden of periprosthetic joint infection of the hip and knee in the United States. J. Arthroplasty. 36 (5), 1484–1489 e3. (2021). https://doi.org/10.1016/j.arth.2020.12.005
Zardi, E. M. & Franceschi, F. Prosthetic joint infection. A relevant public health issue. J. Infect. Public. Health. 13 (12), 1888–1891. https://doi.org/10.1016/j.jiph.2020.09.006 (2020).
Thiele, K. et al. Current failure mechanisms after knee arthroplasty have changed: polyethylene wear is less common in revision surgery. J. Bone Joint Surg. Am. 97 (9), 715–720. https://doi.org/10.2106/JBJS.M.01534 (2015).
Guo, H., Xu, C. & Chen, J. Risk factors for periprosthetic joint infection after primary artificial hip and knee joint replacements. J. Infect. Dev. Ctries. 14 (6), 565–571. https://doi.org/10.3855/jidc.11013 (2020).
Gorišek, N. The Development and implementation of new recommendations for perioperative antibiotic prophylaxis duration in elective primary hip and knee replacement surgeries. J. Clin. Med. 15 (10), 3718. https://doi.org/10.3390/jcm15103718 (2026).
Mulpur, P. et al. Efficacy of intrawound vancomycin in prevention of periprosthetic joint infection after primary total knee arthroplasty: a prospective double-blinded randomized control trial. J. Arthroplasty. 39 (6), 1569–1576. https://doi.org/10.1016/j.arth.2024.01.003 (2024).
Saba, B. V. et al. The one-year infection rates after vancomycin powder and dilute povidone-iodine lavage in high-risk primary total joint arthroplasty: a multicenter randomized controlled trial. J. Arthroplasty. https://doi.org/10.1016/j.arth.2026.04.064 (2026).
Muhlberg, E. et al. Renaissance of vancomycin: approaches for breaking antibiotic resistance in multidrug-resistant bacteria. Can. J. Microbiol. 66 (1), 11–16. https://doi.org/10.1139/cjm-2019-0309 (2020).
McGee, M. A., Howie, D. W., Neale, S. D., Haynes, D. R. & Pearcy, M. J. The role of polyethylene wear in joint replacement failure. Proc. Inst. Mech. Eng. H. 211 (1), 65–72. https://doi.org/10.1243/0954411971534692 (1997).
Boyer, B. et al. Is cross-linked polyethylene an improvement over conventional ultra-high molecular weight polyethylene in total knee arthroplasty? J. Arthroplasty. 33 (3), 908–914. https://doi.org/10.1016/j.arth.2017.10.005 (2018).
Implants for surgery - Wear of total knee-joint prostheses. Part 1: Loading and displacement parameters for wear-testing machines with load control and corresponding environmental conditions for tests, ISO 14243-1:2009 (International Organization for Standardization, 2009).
Halder, A. et al. Influence of limb alignment on mediolateral loading in total knee replacement: in vivo measurements in five patients. J. Bone Joint Surg. Am. 94 (11), 1023–1029. https://doi.org/10.2106/JBJS.K.00927 (2012).
Schroeder, S. et al. Does kinematic alignment increase polyethylene wear compared with mechanically aligned components? A wear simulation study. Clin. Orthop. Relat. Res. 480 (9), 1790–1800. https://doi.org/10.1097/CORR.0000000000002245 (2022).
Dreyer, M. J. et al. Experimental and computational evaluation of knee implant wear and creep under in vivo and ISO boundary conditions. Biomed. Eng. Online. 23 (1), 130. https://doi.org/10.1186/s12938-024-01321-0 (2024).
Micheli, B. R., Wannomae, K. K., Lozynsky, A. J., Christensen, S. D. & Muratoglu, O. K. Knee simulator wear of vitamin E stabilized irradiated ultrahigh molecular weight polyethylene. J. Arthroplasty. 27 (1), 95–104. https://doi.org/10.1016/j.arth.2011.03.006 (2012).
Brandt, J. M., Charron, K. D., MacDonald, S. J. & Medley, J. B. Mass gain behaviour of tibial polyethylene inserts during soak testing. Proc. Inst. Mech. Eng. H. 225 (3), 324–331. https://doi.org/10.1177/2041303310392629 (2011).
Oral, E. & Muratoglu, O. K. Radiation cross-linking in ultra-high molecular weight polyethylene for orthopaedic applications. Nucl. Instrum. Methods Phys. Res. B. 265 (1), 18–22. https://doi.org/10.1016/j.nimb.2007.08.022 (2007).
Heinsberg, L. W. & Weeks, D. E. Post hoc power is not informative. Genet. Epidemiol. 46 (7), 390–394. https://doi.org/10.1002/gepi.22464 (2022).
Peng, Z., Lin, X., Kuang, X., Teng, Z. & Lu, S. The application of topical vancomycin powder for the prevention of surgical site infections in primary total hip and knee arthroplasty: a meta-analysis. Orthop. Traumatol. Surg. Res. 107 (4), 102741. (2021). https://doi.org/10.1016/j.otsr.2020.09.006
Liao, S., Yang, Z., Li, X., Chen, J. & Liu, J. G. Effects of different doses of vancomycin powder in total knee and hip arthroplasty on the periprosthetic joint infection rate: a systematic review and meta-analysis. J. Orthop. Surg. Res. 17 (1), 546. https://doi.org/10.1186/s13018-022-03445-2 (2022).
Funding
This study was supported by implantcast GmbH (Buxtehude, Germany), which supplied the tested 5C® Knee Systems. The sponsor had no role in study design, simulator testing, data collection, data analysis, interpretation of the results, manuscript preparation, or the decision to submit the manuscript for publication.
Author information
Author notes
Celina Lesnicar and Irem Tural contributed equally to this work.
Authors and Affiliations
Department of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU Munich, Marchioninistr. 15, 81377, Munich, Germany
Celina Lesnicar, Irem Tural, Manuel Kistler, Christoph Thorwächter, Richard Zaccaria, Gautier Beckers, Boris M. Holzapfel & Dominic Simon
Authors
- Celina Lesnicar
- Irem Tural
- Manuel Kistler
- Christoph Thorwächter
- Richard Zaccaria
- Gautier Beckers
- Boris M. Holzapfel
- Dominic Simon
Contributions
CL*: data curation, formal analysis, methodology, project administration, writing—original draft, writing – review and editing. IT*: data curation, formal analysis, methodology, writing—original draft, writing—review and editing. MK: methodology, writing—review and editing. CT: methodology, writing—review and editing. RZ: methodology, writing—review and editing. GB: writing—original draft, writing—review and editing. BMH: funding acquisition, writing—review and editing. DS: conceptualization, funding acquisition, methodology, project administration, supervision, writing—original draft, writing—review and editing. *CL and IT contributed equally to this work.
Corresponding author
Correspondence to Dominic Simon.
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lesnicar, C., Tural, I., Kistler, M. et al. Effect of vancomycin-containing lubricant on vitamin E-stabilized UHMWPE wear in an in vitro total knee simulator: a pilot study. Sci Rep 16, 31122 (2026). https://doi.org/10.1038/s41598-026-74531-2
Received: 14 February 2026
Accepted: 29 September 2026
Published: 06 October 2026
Version of record: 06 October 2026
DOI: https://doi.org/10.1038/s41598-026-74531-2