使用结合可见光和近红外光谱数据的无创血糖检测系统
Non-invasive blood glucose detection system using combined visible and near-infrared spectral data
本试点研究评估了使用指尖可见光-近红外光反射非侵入性连续血糖监测(CGM)血糖估计的可行性,共10名成人参与。研究采用随机森林模型,在独立验证中实现了平均绝对误差11.248 mg/dL、均方根误差17.326 mg/dL、相对平均偏差10.379%和R² 0.767。
- Article
- Open access
- Published: 16 November 2026
Scientific Reports (2026) Cite this article
We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.
Abstract
This pilot study evaluated non-invasive estimation of continuous glucose monitoring (CGM)-reported glucose from fingertip visible–near-infrared (VIS–NIR) reflectance in ten adults. Optical measurements (approximately 12 per minute) were retained as separate 70-feature vectors and paired with 1-minute FreeStyle Libre 2 CGM readings checked approximately 10 min afterward using clock-based timing; multiple optical records therefore shared one CGM label. The dataset contained 6,816 optical samples. Seven regression models were compared using sample-level five-fold and subject-independent leave-one-subject-out (LOSO) cross-validation. Under LOSO, Random Forest achieved MAE 11.248 mg/dL, RMSE 17.326 mg/dL, MARD 10.379%, and R² 0.767. In Clarke Error Grid analysis against CGM values, 82.40% and 85.30% of predictions were in Zones A and A + B, respectively. LOSO DNN classification accuracies were 92.74% and 92.43% for 30 and 15 mg/dL categories. These findings support feasibility but indicate inter-subject variability. The manually timed 10-minute offset was approximate and not an individually validated physiological lag correction. No independent capillary or laboratory blood glucose reference was available; thus clinical blood glucose measurement accuracy or safety was not established.
Funding
This work was supported by the Electronics and Telecommunications Research Institute (ETRI) grant funded by the Korean government, and by the Gimhae Biomedical Industry Promotion Agency under the following projects: 26ZD1140, Advancement and Commercialization of Daegu-Gyeongbuk Regional Strategic Industries (Robots, Mobility, AI, Medical, etc.; and 26PD1200, Infrastructure Construction Project of the Untact Medical Service Industry.
Author information
Authors and Affiliations
Daegu-Gyeongbuk Research Center Electronics and Telecommunications Research Institute (ETRI), Daegu, 42994, Republic of Korea
Tae Wuk Bae & Kee Koo Kwon
Electronic Advanced Development Team, AJIN Industrial CO., LTD, Gyeongsan, 38462, South Korea
Byoung Ik Kim
Department of Information and Communication Security, U1 University, Asan, 31415, Republic of Korea
Young Choon Kim
Authors
- Tae Wuk Bae
- Byoung Ik Kim
- Kee Koo Kwon
- Young Choon Kim
Corresponding authors
Correspondence to Tae Wuk Bae or Byoung Ik Kim.
Ethics declarations
Competing interests
Author Byoung Ik Kim is employed by AJIN Industrial Co., Ltd., South Korea. The remaining authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Ethics approval
This study was reviewed and approved by the Institutional Review Board (IRB) of the Electronics and Telecommunications Research Institute (ETRI), Republic of Korea (IRB approval number: N01-202502-01-002). All experimental procedures were performed between February 19, 2025, and April 30, 2025, in accordance with the ethical standards laid down in the 1964 Declaration of Helsinki and its later amendments. A total of ten adult participants (seven men and three women), aged 40–70 years, were enrolled in the study.
Consent to participate
Written informed consent was obtained from all individual participants included in the study, following full disclosure of the study objectives, experimental procedures, potential risks, and data privacy protection measures.
Consent to publish
Not applicable. This manuscript does not contain any identifiable images or personal data of individual participants.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Bae, T.W., Kim, B.I., Kwon, K.K. et al. Non-invasive blood glucose detection system using combined visible and near-infrared spectral data. Sci Rep (2026). https://doi.org/10.1038/s41598-026-80548-4
Received: 12 August 2026
Accepted: 09 November 2026
Published: 16 November 2026
DOI: https://doi.org/10.1038/s41598-026-80548-4
来源:Nature 健康护理主题 · nature.com