智能床移加速器上的一种梯度多中心VMAT策略用于头脊照射

A gradient multi-isocenter VMAT strategy for efficient craniospinal irradiation on an intelligent couch-shifting accelerator

AI 导读

一项研究评估了智能床移加速器上的梯度多中心VMAT策略在头脊照射中的应用,结果显示增加等中心数量提高了剂量调制能力,减少了正常组织的低剂量暴露和治疗时间,同时保持了临床可接受的目标剂量分布。该策略为没有专用调强放疗系统的机构提供了复杂超大照射野放疗的实用选择。

正文

We’re sharing this article early to provide faster access to peer-reviewed, accepted research. It is citable and carries a permanent DOI. This version is subject to further edits and will be replaced automatically by the final Version of Record. All legal disclaimers apply.

Abstract

Delivering high-quality craniospinal irradiation (CSI) remains challenging because of the need to achieve adequate target coverage across ultra-extended volumes while minimizing dose to normal tissue and maintaining clinically practical treatment efficiency. Although helical tomotherapy (HT) provides excellent dose conformity for CSI, prolonged delivery time and increased low-dose exposure may limit its clinical applicability. This study aimed to investigate whether increasing the number of isocenters in a gradient multi-isocenter volumetric modulated arc therapy (VMAT) strategy could enhance modulation capability for extended target volumes and provide a clinically feasible alternative to HT. Sixteen patients who received CSI at The Fourth Hospital of Hebei Medical University between January 2022 and March 2026 were retrospectively selected. For each patient, two VMAT strategies using an intelligent couch-shifting accelerator (TaiChiPro) were generated: a conventional multi-isocenter configuration (Group A) and an increased-isocenter configuration (Group B). HT plans delivered on a Radixact™ system were generated as reference plans (Group C). Dosimetric parameters of the planning target volume (PTV), organs at risk (OARs), low-dose exposure of normal tissue, monitor units (MU), delivery time, and patient-specific quality assurance results were evaluated. All plans met clinical dose constraints. Increasing the number of isocenters improved the dose modulation capability of VMAT for extended target volumes. Compared with the conventional VMAT approach, the increased-isocenter VMAT strategy achieved lower V5, V10, and Dmean values for selected OARs, including the parotid glands and kidneys, as well as reduced low-dose exposure to normal tissue (P < 0.05). Compared with HT, both VMAT strategies substantially reduced delivery time by approximately 78% (1355.98 ± 219.63 s vs. 301.08 ± 27.56 s vs. 302.30 ± 19.02 s, P < 0.001) and decreased low-dose exposure to selected normal tissue. HT maintained advantages in target conformity index (CI), homogeneity index (HI), as well as in reducing Dmax to selected central nervous system structures. However, no significant differences were observed between conventional and increased-isocenter VMAT regarding target coverage. Both VMAT strategies achieved clinically acceptable dose continuity across the target junction regions. Patient-specific quality assurance (QA) further demonstrated reliable delivery accuracy, with gamma passing rates exceeding 97% across all evaluated plans. A gradient multi-isocenter VMAT strategy implemented on an intelligent couch-shifting accelerator offers improved dose modulation for craniospinal irradiation while substantially improving delivery efficiency and reducing low-dose exposure to normal tissue. Increasing the number of isocenters may further improve OARs sparing without substantially increasing treatment time, while maintaining clinically acceptable target dose distributions. This approach provides a practical option for complex ultra-extended field radiotherapy, particularly in institutions without dedicated tomotherapy systems.

Abbreviations

HT:

Helical tomotherapy

CSI:

Craniospinal irradiation

VMAT:

Volumetric modulated arc therapy

PTV:

Planning target volume

OARs:

Organs at risk

MU:

Monitor units

CI:

Conformity index

HI:

Homogeneity index

CTV:

Clinical target volume

PTV-1:

Cranial PTV (the cranial CTV + 3 mm)

PTV-2:

Spinal PTV (the spinal CTV + 5 mm)

PTV-all:

The union of PTV-1 and PTV-2

TPS:

Treatment planning system

FFF:

Flattening filter-free

DVH:

Dose-volume histogram

MC:

Monte Carlo

MLC:

Multi-leaf collimator

Funding

The research was partly supported by Central Guide Local S&T Development Program of Hebei (No. 246Z7722G).

Author information

Author notes

  1. Ruohui Zhang and Wenwen Bai contributed equally to this work.

Authors and Affiliations

  1. Department of Radiation Oncology, The Fourth Hospital of Hebei Medical University, Jiankang Road No.12, Shijiazhuang, 050011, China

    Saimeng Li, Jing Wang, Ning Dai, Xiaohan Li, Yang Zhao, Zifeng Chi, Ruohui Zhang & Wenwen Bai

  2. Department of Gastroenterology, Hebei General Hospital, Shijiazhuang, 050051, China

    Yulan Gao

Authors

  1. Saimeng Li
  2. Yulan Gao
  3. Jing Wang
  4. Ning Dai
  5. Xiaohan Li
  6. Yang Zhao
  7. Zifeng Chi
  8. Ruohui Zhang
  9. Wenwen Bai

Corresponding authors

Correspondence to Ruohui Zhang or Wenwen Bai.

Ethics declarations

Competing interests

The authors declare that they have no competing interests.

Ethics approval and consent to participate

This study was conducted in accordance with the principles of the Declaration of Helsinki. The study was approved by the Medical Ethics Committee of The Fourth Hospital of Hebei Medical University (2026KY215). The requirement for informed consent was waived due to the retrospective nature of the research.

Consent for publication

Not applicable.

Additional information

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, S., Gao, Y., Wang, J. et al. A gradient multi-isocenter VMAT strategy for efficient craniospinal irradiation on an intelligent couch-shifting accelerator. Sci Rep (2026). https://doi.org/10.1038/s41598-026-77634-y

Download citation

  • Received: 14 August 2026

  • Accepted: 21 October 2026

  • Published: 13 November 2026

  • DOI: https://doi.org/10.1038/s41598-026-77634-y

Keywords

来源:Nature 健康护理主题 · nature.com