Retention of graphene-reinforced polymethyl methacrylate crowns bonded to titanium-base abutments: an exploratory in vitro study
该研究通过体外实验评估了不同树脂粘接剂和化学预处理对石墨烯增强聚甲基丙烯酸甲酯冠在钛基台上的粘接力和剩余粘接剂分布的影响。结果显示,树脂粘接剂类型和化学预处理均显著影响粘接力,但无交互作用。研究未发现不同预处理方案间剩余粘接剂分布有显著差异。
Introduction
Titanium-base abutments provide a prefabricated titanium interface for implant-supported computer-aided design and computer-aided manufacturing (CAD-CAM) restorations. They facilitate digital workflows and extraoral bonding, but loss of retention between the restoration and titanium base remains clinically relevant. Retention is influenced by restorative material, cement, abutment geometry, and surface treatment [1,2,3,4,5,6,7,8,9,10].
Graphene-reinforced polymethyl methacrylate (PMMA) is a CAD-CAM polymer proposed for prosthodontic applications. Graphene-based fillers may modify mechanical, biological, antimicrobial, and surface properties relative to conventional PMMA [11,12,13,14,15,16,17,18,19,20]. Its machinability and resilience make it a potential material for interim or definitive restorations, but direct evidence on bonding to titanium-base abutments is limited.
Titanium-base studies have mainly evaluated zirconia, lithium disilicate, hybrid ceramics, resin-matrix ceramics, and conventional or interim PMMA [1,2,3,4,5,6,7,8,9,10, 21,22,23,24,25,26]. Airborne-particle abrasion, primers, and resin cements can alter retention, but findings cannot be directly extrapolated to graphene-reinforced PMMA. Based on these material-specific characteristics, we hypothesized that filler content, polymer matrix, and manufacturing process could influence surface activation and interaction with resin cements.
Therefore, the purpose of this exploratory in vitro study was to evaluate the effects of resin cement type and chemical pretreatment, including their interaction, on maximum pull-out force and residual-cement distribution of graphene-reinforced PMMA CAD-CAM crowns bonded to titanium-base abutments. The null hypotheses were that (1) cement type would not affect maximum pull-out force, (2) chemical pretreatment would not affect maximum pull-out force, (3) there would be no cement × pretreatment interaction, and (4) residual-cement score distributions would not differ among protocols.
Materials and Methods
Study design
This exploratory in vitro study used a 2 × 2 factorial design to evaluate graphene-reinforced polymethyl methacrylate (PMMA) CAD-CAM crowns bonded to titanium-base abutments. The factors were resin cement (Multilink Hybrid Abutment or RelyX Unicem 2) and chemical surface pretreatment (with or without pretreatment), yielding four protocols: Multilink Hybrid Abutment with pretreatment (Group A), RelyX Unicem 2 with pretreatment (Group B), Multilink Hybrid Abutment without pretreatment (Group C), and RelyX Unicem 2 without pretreatment (Group D).
No a priori sample-size calculation was performed because directly comparable data for graphene-reinforced PMMA crowns bonded to titanium-base abutments were unavailable. Six specimens per protocol were included as a pragmatic sample size for this exploratory laboratory investigation, close to the seven specimens per group used in analogous pull-out studies [27,28,29]. The study was not prospectively powered for confirmatory inference; therefore, nonsignificant comparisons were interpreted cautiously and not as evidence of equivalence.
Specimen preparation
Twenty-four graphene-reinforced PMMA crowns were designed using Autodesk Meshmixer, Zirkonzahn Exocad, and Zirkonzahn Nesting software and milled from graphene-reinforced PMMA discs (G-CAM, Graphenano Dental, Valencia, Spain) using Zirkonzahn M5 and M2 milling units. All crowns were generated from the same design file. The cement space was set at 30 µm. The external design simulated an upper premolar crown and included two lateral projections to allow fixation during the pull-out test. The internal design was adapted to receive a titanium-base abutment. The specimen design and pull-out testing protocol were adapted from previous in vitro studies evaluating the retention of CAD-CAM restorations bonded to titanium-base abutments [6, 7]. The specimen preparation and pull-out testing workflow is shown in Fig. 1.
A CAD-CAM nesting of graphene-reinforced polymethyl methacrylate specimens. B Specimen and metallic support assembly. C Surface treatment and cementation. D Specimen mounted in the universal testing machine. E Representative specimens after testing.
Titanium-base abutments (Custom Ti-Base, IPD Dental Group, Barcelona, Spain; ref. IPD/FA-IN-02/3D; 3.5 mm platform) had an original height of 8 mm and were standardized to 5 mm using a cutting guide and a tungsten rotary disc. No additional polishing procedure was performed; the same disc was used to smooth the cut edge and eliminate sharp margins. Each titanium base was screwed onto an implant analogue and torqued according to the manufacturer’s instructions. The analogue was embedded in autopolymerizing acrylic resin within a metallic support designed to be fixed to the lower clamp of the universal testing machine.
Materials
The materials used in the study are summarized in Table 1. Two resin cements were evaluated: Multilink Hybrid Abutment and RelyX Unicem 2. Chemical surface pretreatment consisted of Monobond Plus for the titanium-base surface and SR Connect for the intaglio surface of the graphene-reinforced PMMA crowns.
Surface treatment and cementation protocols
The intaglio surfaces of the graphene-reinforced PMMA crowns and the bonding surfaces of the titanium-base abutments were airborne-particle abraded with 50 µm aluminium oxide particles at 2-3 bar, from a distance of approximately 5 cm for 30 s. Both surfaces were then cleaned with water and oil-free air; the crown intaglio surfaces were additionally cleaned with alcohol and allowed to dry.
After fabrication, specimens were distributed among the four experimental groups (n = 6 per group) according to the cementation protocol. No formal randomization procedure was used. Group A comprised crowns cemented with Multilink Hybrid Abutment after chemical surface pretreatment; Group B, RelyX Unicem 2 after pretreatment; Group C, Multilink Hybrid Abutment without pretreatment; and Group D, RelyX Unicem 2 without pretreatment.
For the chemical surface pretreatment groups, Monobond Plus was applied to the titanium-base surface with a disposable brush, left undisturbed for 60 s, and air-dried. SR Connect was applied to the intaglio surface of the graphene-reinforced PMMA crowns, rubbed for 30 s, air-dried, and light-polymerized for 40 s. Chemical surface pretreatment was performed according to the manufacturers’ instructions and previous protocols used for bonding CAD-CAM restorations to titanium-base abutments [6,7,8, 10].
The resin cements were applied to the internal surface of the crowns using automix syringes according to the manufacturers’ instructions. Before cementation, the screw-access channel was temporarily blocked with cotton to prevent cement ingress; the cotton was removed after cementation. The crowns were seated onto the titanium-base abutments under manual pressure for 5 seconds. For RelyX Unicem 2, excess cement was removed, followed by light-polymerization for 20 s and an additional 6-minute self-curing period. For Multilink Hybrid Abutment, excess cement was removed after 2–3 min, and the specimens were left to self-cure for an additional 7 min. After cementation, specimens were stored dry for 1 week at 20 °C in a laboratory container before pull-out testing; no water-ageing protocol was performed.
Pull-out retention test
Pull-out testing was performed using a universal testing machine (Zwick/Roell Z100, Ulm, Germany). Each specimen was positioned so that the titanium-base/analogue assembly was fixed to the lower clamp and the lateral projections of the crown were fixed to the upper clamp (Fig. 1D). A preload of 0.1 MPa was applied and tensile loading was performed along the long axis at a crosshead speed of 1 mm/min until debonding or fracture. The laboratory software originally expressed the maximum value as stress using nominal specimen dimensions of 3 × 3 mm (9 mm²). Because this area did not represent the actual titanium-base/crown bonding surface, the revised primary outcome is maximum pull-out force (N), recovered by multiplying the recorded stress value by 9 mm².
Residual-cement assessment
After testing, the debonded surfaces were visually inspected to assess the location of residual cement (Fig. 1E). Residual-cement distribution was assessed by a single examiner, who evaluated all specimens together using a predefined five-point visual code; no formal examiner-blinding procedure was implemented. The code was adapted from previous studies: 6, 7, 10 score 1, cement remained entirely on the titanium-base abutment; score 2, more than 50% remained on the titanium-base abutment; score 3, cement was distributed approximately equally between crown and titanium base; score 4, more than 50% remained on the crown; and score 5, cement remained entirely on the crown.
Statistical analysis
Statistical analysis was performed with jamovi, version 2.2 (The jamovi project, Sydney, Australia). Means, standard deviations, standard errors, minimums, maximums, and 95% confidence intervals were calculated. Homogeneity of variance was assessed using Levene’s test. Because variances were heterogeneous (P = 0.021), protocol-level comparisons used Welch one-way analysis of variance followed by Games-Howell pairwise tests. To reflect the 2 × 2 factorial design, a linear model including cement type, chemical pretreatment, and the cement × pretreatment interaction was also fitted. Conventional factorial F tests and partial eta-squared (ηp²) were reported as effect-size estimates; because of heteroscedasticity, the interaction was additionally evaluated using HC3 heteroscedasticity-robust standard errors. Residual-cement score distributions were evaluated using Pearson chi-square and Fisher’s exact tests. The significance level was α = 0.05.
Use of generative AI and AI-assisted technologies
During revision of this manuscript, the authors used ChatGPT (OpenAI) to assist with language editing, organisation, and clarity. The authors reviewed and verified the resulting text, made all necessary revisions, and take full responsibility for the content of the manuscript.
Ethical approval
Not applicable. This exploratory in vitro laboratory study did not involve human participants, human tissues, or animals; therefore, ethics committee approval was not required.
Results
All 24 specimens were included in the analysis. Descriptive maximum pull-out force values are shown in Table 2, the group means are displayed in Fig. 2, and residual-cement scores are presented in Table 3. Individual raw values are provided in the accompanying underlying-data file.
Error bars indicate standard deviations. Groups sharing at least one letter were not significantly different in the exploratory Games-Howell comparisons (P < 0.05). MHA Multilink Hybrid Abutment, RU2 RelyX Unicem 2.
Mean maximum pull-out force was 452.9 ± 74.0 N (95% CI, 375.2–530.5 N) in Group A, 688.7 ± 185.1 N (95% CI, 494.4–882.9 N) in Group B, 400.7 ± 94.7 N (95% CI, 301.3–500.0 N) in Group C, and 366.8 ± 82.4 N (95% CI, 280.3–453.2 N) in Group D. Factorial analysis showed a significant cement × pretreatment interaction (F = 7.86; P = 0.011; ηp² = 0.282). The interaction remained significant with HC3 robust inference (interaction coefficient, 269.7 N; 95% CI, 49.9–489.5 N; P = 0.019). The estimated pretreatment-associated increase was 52.2 N for Multilink Hybrid Abutment and 321.9 N for RelyX Unicem 2. Conventional factorial main effects were cement: F = 4.41, P = 0.049, ηp² = 0.181; and pretreatment: F = 15.13, P < 0.001, ηp² = 0.431; given the significant interaction, these main effects were not interpreted in isolation.
As an exploratory protocol-level analysis, Welch one-way analysis of variance also detected differences among the four protocols (F = 4.86; df1 = 3; df2 = 10.8; P = 0.022). Games-Howell comparisons showed that Group B exceeded Group C (mean difference, 288.0 N; P = 0.042) and Group D (mean difference, 321.9 N; P = 0.025). Group A did not differ significantly from Group B (P = 0.090), Group C (P = 0.718), or Group D (P = 0.287), and Groups C and D did not differ significantly (P = 0.909). These nonsignificant comparisons were not interpreted as evidence of equivalence.
Residual-cement score distributions differed among protocols (χ² = 28.3; df = 9; P < 0.001; Fisher exact P < 0.001). In Group A, score 2 was most frequent (4/6 specimens, 66.7%), with one score 3 and one score 4. In Group B, scores 2 and 3 were equally frequent (3/6 each, 50.0%). Every specimen in Groups C and D had score 1, indicating failure at the crown-cement interface.
Discussion
This exploratory in vitro study evaluated a 2 × 2 cementation design for graphene-reinforced PMMA crowns bonded to titanium-base abutments. The principal finding was a significant cement × pretreatment interaction, indicating that the association between chemical pretreatment and maximum pull-out force depended on the resin cement used. Residual-cement score distributions also differed among protocols; therefore, the null hypotheses of no interaction and no difference in residual-cement distribution were rejected.
The estimated pretreatment-associated increase was modest for Multilink Hybrid Abutment (52.2 N) but substantially larger for RelyX Unicem 2 (321.9 N). The HC3 robust analysis supported the interaction despite heterogeneous variances. Nevertheless, the small sample size produced wide confidence intervals, particularly for pretreated RelyX Unicem 2, and the estimates should be considered exploratory. Nonsignificant contrasts, including Multilink Hybrid Abutment with versus without pretreatment, should not be interpreted as demonstrating equivalence.
Residual-cement distribution provided complementary information about the bonded interface. Group A most frequently showed score 2, whereas Group B showed scores 2 and 3 with equal frequency. In contrast, all non-pretreated specimens had score 1, with cement remaining entirely on the titanium base. This pattern is consistent with failure predominantly at the crown-cement interface in the absence of chemical pretreatment, whereas pretreated groups showed a more mixed distribution. Because scoring was visual and performed by a single non-blinded examiner, the residual-cement findings should be interpreted as descriptive support rather than mechanistic proof.
All bonding surfaces were airborne-particle abraded with 50-µm aluminium oxide before cementation. Airborne-particle abrasion can increase surface roughness and micromechanical retention, but the observed failure patterns indicate that abrasion alone did not establish a stable interaction between the tested graphene-reinforced PMMA and either resin cement [4, 6,7,8,9,10, 21]. SR Connect may promote interaction with the PMMA-based matrix through methacrylate groups, while Monobond Plus contains functional monomers intended for bonding to titanium. This mechanistic interpretation remains provisional because surface chemistry and morphology were not directly characterised.
Previous titanium-base studies have mainly investigated zirconia, lithium disilicate, hybrid ceramics, resin-matrix ceramics and conventional or interim PMMA [1,2,3,4,5,6,7,8,9,10, 21,22,23,24,25,26]. The present study was based on the hypothesis that filler content, polymer matrix, and manufacturing process could influence surface activation and cement interaction in graphene-reinforced PMMA. The observed cement-dependent interaction supports the need for material- and protocol-specific evaluation rather than direct extrapolation from ceramic or conventional PMMA systems.
Graphene-containing PMMA materials have been proposed for prosthodontic use because graphene-based fillers may alter mechanical, antimicrobial, optical, and surface properties [11,12,13,14,15,16,17,18,19,20, 30]. Evidence remains heterogeneous, and these properties may vary with graphene concentration, dispersion, and manufacturing conditions. Bonding protocols should therefore be tested for the specific commercial material rather than assumed to behave like conventional PMMA or ceramic CAD-CAM materials.
Clinically, extraoral bonding to titanium bases is central to many digital implant workflows, and loss of retention remains a relevant complication [31]. The present findings suggest that, under the immediate laboratory conditions tested, chemical pretreatment may have a larger effect when RelyX Unicem 2 is used than when Multilink Hybrid Abutment is used [32]. They do not establish a universal clinical recommendation because only one graphene-reinforced PMMA, one titanium-based system, two cements, and one pretreatment combination were evaluated without thermomechanical ageing.
Study limitations and future research
The study has several limitations. It was exploratory, included only six specimens per protocol, and was not prospectively powered for confirmatory inference. The small sample limits precision and statistical power, and nonsignificant comparisons should not be interpreted as evidence of equivalence. Crown seating was performed under manual pressure for 5 seconds rather than a standardized load, which may have introduced variation in cement thickness. Specimens were distributed among groups without a formal randomization procedure. Residual-cement assessment was visual and performed by a single examiner without formal blinding.
The specimens were stored for 1 week at 20 °C but were not thermocycled, water-aged or mechanically fatigued, and the pull-out test applied a single axial tensile load rather than the complex multidirectional and cyclic forces present intraorally [21,22,23,24,25]. Failure patterns were assessed visually rather than by microscopy or surface analysis. Only one graphene-reinforced PMMA, one titanium-base system, two resin cements, and one primer combination were tested; the results cannot be generalized beyond these materials and protocols.
Future studies should use larger prospectively justified samples, standardized cementation pressure, formal random allocation and blinded failure assessment, thermomechanical ageing, and complementary surface characterization. Testing additional cements, primers and titanium-base geometries would help distinguish substrate, cement and interaction effects [33,34,35. Clinical studies would ultimately be required to determine whether the observed differences translate into fewer debonding events.
Conclusions
Within the limitations of this exploratory in vitro study, the effect of chemical pretreatment on maximum pull-out force depended on the resin cement used, as demonstrated by a significant cement × pretreatment interaction.
Pretreatment was associated with a substantially larger increase in maximum pull-out force for RelyX Unicem 2 than for Multilink Hybrid Abutment and with a shift in residual-cement distribution away from uniform failure at the crown-cement interface. These findings are material- and protocol-specific and require confirmation in larger, prospectively powered studies with ageing.
Data availability
The underlying data supporting the findings of this study are provided in the accompanying data file and are also available from the corresponding author upon reasonable request.
References
Silva CEP, Soares S, Machado CM, Bergamo ETP, Coelho PG, Witek L, et al. Effect of CAD/CAM abutment height and cement type on the retention of zirconia crowns. Implant Dent. 2018;27:582–7.
Lopes ACO, Machado CM, Bonjardim LR, Bergamo ETP, Ramalho IS, Witek L, et al. The effect of CAD/CAM crown material and cement type on retention to implant abutments. J Prosthodont. 2019;28:e552–6.
Zahoui A, Bergamo ETP, Marun M, Silva KP, Coelho PG, Bonfante EA. Cementation protocol for bonding zirconia crowns to titanium base CAD/CAM abutments. Int J Prosthodont. 2020;33:527–35.
Bergamo ETP, Zahoui A, Ikejiri LLA, Marun M, da Silva KP, Coelho PG, et al. Retention of zirconia crowns to Ti-base abutments: effect of luting protocol, abutment treatment and autoclave sterilization. J Prosthodont Res. 2021;65:171–5.
Moreno ALDM, dos Santos DM, Bertoz APDM, Goiato MC. Abutment on titanium-base hybrid implant: a literature review. Eur J Dent. 2023;17:261–9.
Pitta J, Burkhardt F, Mekki M, Fehmer V, Mojon P, Sailer I. Effect of airborne-particle abrasion of a titanium base abutment on the stability of the bonded interface and retention forces of crowns after artificial aging. J Prosthet Dent. 2021;126:214–21.
Pitta J, Bijelic-Donova J, Burkhardt F, Fehmer V, Närhi T, Sailer I. Temporary implant-supported single crowns using titanium base abutments: an in vitro study on bonding stability and pull-out forces. Int J Prosthodont. 2020;33:546–52.
Kemarly K, Arnason S, Parke A, Lien W, Vandewalle K. Effect of various surface treatments on Ti-base coping retention. Oper Dent. 2020;45:426–34.
Rödiger M, Kloß J, Gersdorff N, Bürgers R, Rinke S. Removal forces of adhesively and self-adhesively luted implant-supported zirconia copings depend on abutment geometry. J Mech Behav Biomed Mater. 2018;87:119–23.
Filokyprou T, Kesterke MJ, Liu X, Cho SH, Revilla-León M. Effect of different surface treatments on the retention force of additively manufactured interim implant-supported crowns. J Prosthodont. 2024;33:899–907.
Li X, Liang X, Wang Y, Wang D, Teng M, Xu H, et al. Graphene-based nanomaterials for dental applications: principles, current advances, and future outlook. Front Bioeng Biotechnol. 2022;10:804201.
Tahriri M, Del Monico M, Moghanian A, Tavakkoli Yaraki M, Torres R, Yadegari A, et al. Graphene and its derivatives: opportunities and challenges in dentistry. Mater Sci Eng C Mater Biol Appl. 2019;102:171–85.
Bacali C, Badea M, Moldovan M, Sarosi C, Nastase V, Baldea I, et al. The influence of graphene in improvement of physico-mechanical properties in PMMA denture base resins. Mater (Basel). 2019;12:2335.
Bacali C, Baldea I, Moldovan M, Carpa R, Olteanu DE, Filip GA, et al. Flexural strength, biocompatibility, and antimicrobial activity of a polymethyl methacrylate denture resin enhanced with graphene and silver nanoparticles. Clin Oral Investig. 2020;24:2713–25.
Lee JH, Jo JK, Kim DA, Patel KD, Kim HW, Lee HH. Nano-graphene oxide incorporated into PMMA resin to prevent microbial adhesion. Dent Mater. 2018;34:e63–72.
Di Carlo S, De Angelis F, Brauner E, Pranno N, Tassi G, Senatore M, et al. Flexural strength and elastic modulus evaluation of structures made by conventional PMMA and PMMA reinforced with graphene. Eur Rev Med Pharm Sci. 2020;24:5201–8.
Selva-Otaolaurruchi E, Fernández-Estevan L, Solá-Ruiz M, García-Sala-Bonmati F, Selva-Ribera I, Agustín-Panadero R. Graphene-doped polymethyl methacrylate (PMMA) as a new restorative material in implant-prosthetics: in vitro analysis of resistance to mechanical fatigue. J Clin Med. 2023;12:1269.
Agarwalla SV, Malhotra R, Rosa V. Translucency, hardness and strength parameters of PMMA resin containing graphene-like material for CAD/CAM restorations. J Mech Behav Biomed Mater. 2019;100:103388.
Azevedo L, Antonaya-Martin JL, Molinero-Mourelle P, del Río-Highsmith J. Improving PMMA resin using graphene oxide for a definitive prosthodontic rehabilitation: a clinical report. J Clin Exp Dent. 2019;11:e670–4.
Serrano-Belmonte I, Torné-Durán S, San Nicolás-Sánchez J, Pérez-Fernández V, Martínez-Cánovas A. Comparison of color stability between PMMA and graphene-reinforced PMMA after being subjected to pH changes and coffee. Dent J (Basel). 2026;14:319.
Shadid RM. Retention of CAD-CAM implant-supported ceramic restorations luted to titanium bases: a systematic review of in-vitro studies. Clin Cosmet Investig Dent. 2025;17:305–33.
Shadid RM. How much is stable the bonding of CAD-CAM implant-supported all-ceramic restorations to titanium bases clinically? A systematic review. Clin Cosmet Investig Dent. 2025;17:181–200.
Stratos A, Niakou A, Kamalakidis S, Pissiotis A, Tortopidis D, Michalakis K. Effect of cement, surface treatment, and mechanical aging on the retention of zirconia and Co-Cr restorations on titanium base abutments: an in vitro study. J Prosthet Dent. 2026;135:e123–e131. https://doi.org/10.1016/j.prosdent.2025.12.029.
Freitas MIM, Ruggiero MM, Marcello-Machado RM, Vargas-Moreno VF, dos Reis AC, Del Bel Cury AA. Resin matrix ceramic crowns cemented on titanium bases: effect of surface treatments and stress distribution. Braz Dent J. 2025;36:e236140.
Karasan D, Pitta J, Zarauz C, Strasding M, Liu X, Fehmer V, et al. The influence of titanium-base abutment geometry and height on mechanical stability of implant-supported single crowns. Clin Oral Implants Res. 2024;35:1033–41.
Falahchai M, Neshandar Asli H, Daneshvar M, Hendi A. Effect of different surface treatments on retention and fracture strength of zirconia crowns over Ti-base zirconia abutments. Heliyon. 2024;10:e40164.
Gehrke P, Alius J, Fischer C, Erdelt KJ, Beuer F. Retentive strength of two-piece CAD/CAM zirconia implant abutments. Clin Implant Dent Relat Res. 2014;16:920–5.
Alseddiek A, Al-Zordk W, Attia A. Retention of hybrid-abutment-crowns with offset implant placement: influence of crown materials and Ti-base height. BMC Oral Health. 2023;23:784.
Ibrahim EA, Khamis MM, Ezzelarab S, Abdelhamid AM. Retention of zirconia crowns to titanium bases with straight versus angled screw access channels: an in vitro study. BMC Oral Health. 2023;23:458.
Beleidy M, Ziada A. Wear resistance and surface roughness assessment of CAD/CAM graphene-reinforced PMMA versus conventional PMMA crowns. BMC Oral Health. 2025;25:1821.
Prause E, Hey J, Beuer F, Rosentritt M. Factors influencing retention of resin-based luting systems on implants: a systematic review. Dent Rev. 2022;2:100056.
Fouquet V, Dantagnan CA, Abdel-Gawad S, Dursun E, Attal JP, François P. In vitro shear bond strength over zirconia and titanium alloy and degree of conversion of extraoral compared to intraoral self-adhesive resin cements. BDJ Open. 2023;9:54.
ElMalah NS, Ibrahim Y, Mostafa D. Evaluation of 3D printed nano-modified resin shear bond strength on titanium surfaces: an in vitro study. BMC Oral Health. 2025;25:806.
Donmez MB, Çakmak G, Yılmaz D, Schimmel M, Abou-Ayash S, Yilmaz B, et al. Bond strength of additively manufactured composite resins to dentin and titanium when bonded with dual-polymerizing resin cements. J Prosthet Dent. 2024;132:1067.e1–8.
Küçükekenci AS, Dönmez MB, Dede DÖ, Çakmak G, Yilmaz B. Bond strength of recently introduced computer-aided design and computer-aided manufacturing resin-based crown materials to polyetheretherketone and titanium. J Prosthet Dent. 2024;132:1066.e1–8.
Acknowledgements
The authors thank Graphenano Dental for providing the graphene-reinforced PMMA material, IPD Dental Group for providing the titanium-base abutments, and Laboratorio Mayos for technical support during specimen preparation and laboratory procedures.
Funding
No financial funding was received for this research. Graphenano Dental provided the graphene-reinforced polymethyl methacrylate material and IPD Dental Group provided the titanium-base abutments, as described in the Acknowledgements.
Author information
Authors and Affiliations
Master’s Degree in Oral and Maxillofacial Rehabilitation and Prosthodontics, Faculty of Dentistry, University of Barcelona, Barcelona, Spain
Blanca Prats Sisquella & Nikol A. Peeva
Department of Odonto-Stomatology, Faculty of Medicine and Health Sciences, University of Barcelona, Barcelona, Spain
Sergi Torné-Durán
Odontological and maxillofacial pathology and therapeutics (509), Bellvitge Biomedical Research Institute (IDIBELL), L’Hospitalet de Llobregat, Spain
Sergi Torné-Durán
Full Professor of Preventive Medicine (Biostatistics), Faculty of Medicine, University of Barcelona; Epidemiology and Statistics Unit, Fundación Clínica, Barcelona, Spain; Director, Department of Public Health, Faculty of Medicine, University of Barcelona, Barcelona, Spain
Carlos Ascaso-Terren
Department of Odonto-Stomatology, University of Barcelona; Director, Master’s Degree in Oral and Maxillofacial Rehabilitation and Prosthodontics, Faculty of Dentistry, University of Barcelona, Barcelona, Spain
Tomás Escuin-Henar
Authors
- Blanca Prats Sisquella
- Nikol A. Peeva
- Sergi Torné-Durán
- Carlos Ascaso-Terren
- Tomás Escuin-Henar
Contributions
Conceptualization: BP, NAP, and TE-H; methodology: BP, NAP, CA-T, and TE-H; investigation: BP and NAP; specimen preparation and laboratory procedures: BP, NAP, and TE-H; statistical analysis: CA-T; data interpretation: ST-D, CA-T, and TE-H; writing-original draft: BP, NAP, and ST-D; writing-review and editing: ST-D, CA-T, and TE-H; supervision and project administration: TE-H. All authors read and approved the manuscript.
Corresponding author
Correspondence to Sergi Torné-Durán.
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Prats Sisquella, B., A. Peeva, N., Torné-Durán, S. et al. Retention of graphene-reinforced polymethyl methacrylate crowns bonded to titanium-base abutments: an exploratory in vitro study. BDJ Open 12, 104 (2026). https://doi.org/10.1038/s41405-026-00510-4
Received: 31 July 2026
Revised: 02 October 2026
Accepted: 07 October 2026
Published: 02 November 2026
Version of record: 02 November 2026
DOI: https://doi.org/10.1038/s41405-026-00510-4
来源:Nature 健康护理主题 · nature.com